Troubleshooting
Problem
Probable Fault
Solution
No live feed coming from
camera
No power to SeeSnake
Check power is correctly plugged in.
Connection faulty
Check alignment and pins of connection from CS6 to the
SeeSnake Reel.
18 V Li-Ion battery low
Recharge battery. Switch to AC power.
Battery warning appears
18 V Li-Ion battery low
Recharge battery. Switch to AC power.
Count accuracy unreliable
Integrated counter settings incor-
rect for reel or cable
Verify the settings are correct for the SeeSnake cable
length, cable diameter, and SeeSnake reel.
Counting from the wrong ze-
ro-point
Confirm measurement from the intended zero-point.
Reset zero-point using the Zero key .
Integrated counter low battery
warning appears on screen
Integrated counter battery dead
or near-dead
Replace 3 V battery in the integrated counter.
The “+” symbol appears after
the on-screen distance mea-
surement
Physical cable measurement has
exceeded the cable parameter in
settings.
Verify the actual length of installed cable and reset the
SeeSnake reel and cable settings according to the reel’s
operator’s manual.
SOS blinking on Power Key
LED
No video signal
Check SeeSnake system cable connection.