background image

4

 

 

c. 

To clean eyepiece lenses, do not remove from eyepiece tube. Clean only the outer lens surface.  

 

 

 

Breathe on lens to dampen surface, then wipe with lens paper or use a cotton swab 

 

 

 

moistened with distilled water. Wipe lenses with a circular motion, applying as little pressure as  

 

 

 

possible. Avoid wiping dry lens surface as lenses are scratched easily.  If excessive dirt or grease  

 

 

 

gets on lens surfaces, a small amount of Windex can be used on a cotton swab or lens tissue. To  

 

 

 

clean objective lenses, do not remove objectives from microscope.  Clean front lens element only,  

 

 

 

following same procedure.

 

2. 

Mechanical Maintenance

 

 

The only mechanical adjustment the microscope might require is the tension of the focusing mechanism.  

 

 

This has been adjusted at the factory, but over the course of time it may loosen and cause the head of the  

 

 

microscope to slip downward on the focusing block.

 

 

The tension adjustment collar is located between arm and focus knob on left side of microscope when the  

 

 

stage is facing towards you.  With a jewelers type screwdriver, loosen slotted set screw located on knurled  

 

 

surface of the tension adjustment collar. Turn collar clockwise to tighten tension, counter-clockwise to   

 

 

loosen tension.  After adjusting, tighten set-screw to lock collar into place.

 

 

NOTE:  It is recommended that you leave the tension as loose as possible for ease of focusing, yet not so  

 

 

loose that it permits the head of the microscope to drift downward from its own weight and cause the   

 

 

microscope to “drift” out of focus.

TROUBLESHOOTING

 

 

 

 

Problem 

 

 

Reason for Problem  

 

 

Solution

Light Fails to Operate

Outlet inoperative

AC Power Cord not Connected

Lamp Burned Out

Fuse Blown

Have qualified service technician repair 

outlet

Plug into outlet 

Replace Lamp

Replace Fuse

Adjust tension control

Clean Objective Lenses

Clean Eyepiece Lenses

Clean Eyepiece Lenses

Image does not remain in 

focus

Head of microscope drops from its own weight

Objective lenses dirty

Eyepiece lens dirty

Eyepiece lens dirty

Poor Resolution

(Image not sharp)

Spots in field of View

***Spots in field of view can also result from dirt on inside of eyepiece.  It is recommended that you have a 

service technician clean inside of lens