Guidance and manufacturer’s declaration - electromagnetic immunity

The device is intended for use in the electromagnetic environment specified below. 
The customer or the user  should assure that it is used in such an environment.

Immunity test

IEC 60601 test 
level

Compliance 
level

Electromagnetic 
environment guidance

Electrostatic 
discharge  
(ESD) IEC 
61000-4-2

±6 kV contact

±6 kV contact

Floors should be wood, 
concrete or ceramic tile. 
If floors are covered with 
synthetic material, the 
relative humidity should 
be at least 30 %.

±8 kV air

±8 kV air

Electrical fast 
transient/burst 
IEC 61000-4-4

±2 kV for power 
supply

±2 kV for 
power supply

Main power quality should 
be that of a commercial or  
hospital environment. If 
the user of the device 
requires continued 
operation during power 
mains interruptions, it is 
needed that the device 
be powered from an 
uninterruptible power 
supply

Surge IEC 
61000-4-5

±1 kV line(s) to 
line(s)

±1 kV line(s) to 
line(s)

Voltage dips, 
short interrup-
tions and volt-
age variations 
on power supply 
input lines IEC 
61000-4-11

<5% UT (>95% 
dip in UT) for 0.5 
cycle

<5% UT (>95% 
dip in UT) for 
0.5 cycle

40% UT (60% 
dip in UT) for 5 
Cycles

40% UT (60% 
dip in UT) for 5 
Cycles

70% UT (30% 
dip in UT) for 25 
Cycles

70% UT (30% 
dip in UT) for 
25 Cycles

<5% UT (>95% 
dip in UT) for 
5 sec

<5% UT (>95% 
dip in UT) for 
5 sec

NOTE UT is the a.c. mains voltage prior to application of the test level.

Summary of Contents for US 1000 3RD EDITION

Page 1: ...INSTRUCTION MANUAL US 1000 3RD EDITION...

Page 2: ...r notice Amendments may however be published in new editions of this manual All Right Reserved Rev V2 2019 United States Federal Law restricts this device to sale by or on the order of a physician or...

Page 3: ...NS _______________________________________ 9 4 1 Accessories 4 2 Technical information 5 INSTRUCTIONS FOR USE ________________________________ 11 5 1 Function Test of Ultrasonic Action 6 MAINTENANCE A...

Page 4: ...US 1000 is a Therapeutic ultrasound device that generates pulsed high frequency sound waves 1MHz that are transferred to a specific body are via a sound headed probe The Pulsed Sound Waves travel dee...

Page 5: ...DO NOT use for patients with demand type cardiac pacemakers 7 DO NOT use on someone who is pregnant 8 DO NOT use over ischemic tissues in patients with vascular disease where the blood supply would b...

Page 6: ...device away from children 6 DO NOT use over an area of the spinal cord following laminectomy i e when major covering tissues have been removed 7 Avoid bony prominences 8 When using ultrasound keep th...

Page 7: ...tor or any other metallic or electronic device should not use this device without first consulting a doctor 9 Patients with the following diseases or symptoms should not use the device Pregnancy or Me...

Page 8: ...r the patient with bleeding physique DO NOT use US 1000 WARNING The device complies completely with all parts of 21 CFR 1050 10 of the performance standard for sonic infrasonic and ultrasonic radiatio...

Page 9: ...LED INTENSITY ON OFF MAIN BODY INDICATOR HI MED LO BUTTON TREATMENT HEAD MAIN BODY 4 SPECIFICATIONS 4 1 Accessories NO DESCRIPTION QTY 1 Ultrasound Device DU1025 1 2 Wall Adapter DU1025X 1 3 Therason...

Page 10: ...Humidity Dimensions 6 80 x 2 25 x 1 65 inches L W H Weight 4 25 ounces Frequency 1MHz 10 Type Non focusing Pulse Width 8 0 5ms 10 14 1ms 10 29 2ms 10 Repetition Rate 150 Hz 10 Temporal Maximum Power 4...

Page 11: ...s and will void warranty Apply Transmission Gel Wash the area to be treated so that it is free of oil and dirt Apply a generous layer of ultrasound transmission gel on the treatment area The gel acts...

Page 12: ...er shut down press the On Off button and hold down for 4 seconds and the device will turn off 5 1 Function Test of Ultrasonic Action Place the probe horizontally then apply several water drops on the...

Page 13: ...ected and tested by a Service Technician certified by an Authorized Service Center Cleaning the Applicator The applicator should be regularly inspected for damage e g hairline cracks which could allow...

Page 14: ...cal distributor or the manufacturer as shown on the label PROBLEM CHECK POINTS POSSIBLE SOLUTION LED light does not turn on 1 The plug adapter is not inserted into the socket properly 2 The DC plug of...

Page 15: ...interferences as well as maximum levels of electromagnetic emissions for medical devices Medical devices manufactured by Compass Health Brands Corp conform to this IEC50501 1 2 2007 standard for both...

Page 16: ...up 1 The device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11...

Page 17: ...000 4 4 2 kV for power supply 2 kV for power supply Main power quality should be that of a commercial or hospital environment If the user of the device requires continued operation during power mains...

Page 18: ...frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF lEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m NOTE I At 80 MHz and 800 MHz the separati...

Page 19: ...e according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10 3 69 3 69 7 38 100 11 67 11 67 23 33 For tra...

Page 20: ...sposed of with household waste after their useful life Help us to protect the environment and save resources and take this device to the ap propriate collection points Please contact the organization...

Page 21: ...manufacturer to the consumer or during transport to the retailer Accessories which are subject to normal wear and tear Liability for direct or indirect consequential losses caused by the unit is exclu...

Page 22: ...and return freight charges will apply All returns of dropshipped items are subject to a restocking fee as well as inbound and outbound freight charges 7 Returns will not be accepted on items that are...

Page 23: ...23...

Page 24: ...Manufactured for 3289 19 06 A 2019 Compass Health Brands Corp...

Reviews: