Richmar APT Performer Operation Manual Download Page 25

GUIDANCE AND MANUFACTURER’S DECLARATION –   

ELECTROMAGNETIC IMMUNITY

The APT Performer is intended for use in the electromagnetic environment specified 

below. The customer or the user of APT Performer should assure that it is used in such an 

environment.

Immunity test  IEC 60601 test level 

Compliance level 

Electromagnetic environment – 

guidance

Conducted RF
IEC 61000-4-6

3 Vrms

150 kHz - 80 MHz

6 V rms in ISM bands 

between 150 kHz 

and 80 MHz

80 % AM at 1 kHz

3 Vrms

150 kHz - 80 MHz

6 V rms in ISM bands 

between 150 kHz and 

80 MHz

80 % AM at 1 kHz

N/A

Radiated 

RF

IEC 61000-

4-3

3 V/m

80 MHz - 2.7 GHz

80 % AM at 1 kHz

3 V/m

80 MHz - 2.7 GHz

80 % AM at 1 kHz

N/A

Proximity 

fields from 

RF wireless 

communica-

tions 

equipment

IEC 61000-

4-3

See table: Test specifications for 

ENCLOSURE PORT IMMUNITY to RF 

wireless communications equipment.

N/A

23

APPENDIX A - EMC TABLES

Summary of Contents for APT Performer

Page 1: ...OPERATION MANUAL APT Performer...

Page 2: ...A PHYSICIAN LICENSED PRACTITIONER OR QUALIFIED MEDICAL PERSONNEL IN A CLINICAL ENVIRONMENT AND HAVE THE BASIC PHYSICAL COGNITIVE PREREQUISITES SUCH AS VISION HEARING AND LITERACY AS WELL AS BASIC FUNC...

Page 3: ...ights and you may also have other rights which vary from location to location Richmar does not authorize any person or representative to create for it any other obligation or liability in connection w...

Page 4: ...plains possible safety infractions that could have the potential to cause minor to moderate injury and or damage to equipment WARNING Text under a WARNING heading explains possible safety infractions...

Page 5: ...ws of all adjustable parts of the device handlebar supporting module calf supports are tightened and intact Should they become loose during training stop the training immediately and tighten the screw...

Page 6: ...erate the mains plug Ensure that the power cord is easily detachable in an emergency situation Take care not to allow water to enter the device Keep cables and electric components away from water and...

Page 7: ...k fire or personal injury If a malfunction occurs discontinue use immediately disconnect the power cord from the outlet and contact tech support NEVER perform unauthorized service work All service wor...

Page 8: ...1 Ligaments knees or joint ruptures 2 Post osteoporosis 3 Crippled limbs 4 A person with severe spasticity or stiffness in the limbs 5 Acute thrombosis 6 Pregnancy 7 Patients where vital signs are uns...

Page 9: ...60 r min Resistance Level 0 24 Spasm Level 1 5 Spasm Relief Rate 1 5 Timer duration time 1 60 min 7 ELECTRICAL Power Supply AC100 240V 50 60Hz Rated Power 160 VA Mode of Operation Continuous System of...

Page 10: ...ULT ROTATION FWD REV FWD SCREEN BRIGHTNESS 1 10 1 or 2 5 VOLUME 1 3 1 2 DEVICE DESCRIPTION TECHNICAL SPECIFICATIONS CONTINUED BLUETOOTH Bluetooth Version 4 0 Frequency Range 2 400 GHz 2 483 GHz Modula...

Page 11: ...emove the device from packaging pulling upwards Remove the remaining foam material around all accessories INSPECTIONS Immediately upon unpacking the device perform the following steps 1 Verify the del...

Page 12: ...nt is training from a wheelchair the wheelchair should be secured using the anti tipper wheelchair hooks with patients wheels in locked position to avoid any movement during training If the patient is...

Page 13: ...of images below illustrating how to secure the patient s arms in the straps DEFINITIONS OF TRAINING PROGRAMS Strength Training Program Strength training is designed to encourage the patient to use mor...

Page 14: ...t set correctly this could cause records to be lost or stored inaccurately Press the EDIT button to set the Date and Time then press OK button Set the value according to the local time before first us...

Page 15: ...RPMs If CANCEL is pressed the previous interface will appear on the screen to lower the RPMs below 30 GAMEPLAY Each game interface shows tips about how to play each game Below is one example Game Sta...

Page 16: ...ing please press the PAUSE button Adjust the parameters then press RESUME button to finish training Strength Training The Strength Training is designed to increase resistance when the big fish approac...

Page 17: ...gram designed to encourage the patient to maintain an even balance between their left and right side The Symmetry training parameters and game are as shown below After the training parameters have bee...

Page 18: ...the number of bugs that appear on the left and right of the screen Press the arrows to move the ratio to the right or left Read the game tips and tell the patient how to train then press OK button to...

Page 19: ...Section for more details STORING AND RECALLING FAVORITES Storing Favorites To store a training session as a favorite to use for future trainings or on other patients press FAVORITES button from the T...

Page 20: ...handlebar to break the motor circuit The trainer motor will stop immediately Prompt appears on the screen 2 Release emergency stop Reset the emergency stop button by pulling out the red button Then t...

Page 21: ...ories must be checked at regular intervals INTERVAL ROUTINE MAINTENANCE WEEKLY OBVIOUS DAMAGE Visually inspect components for damage that could cause problems Check the power cord to ensure there is n...

Page 22: ...ot adjustment knobs are not completely tightened 1 Place the trainer in a smooth position for training 2 Tighten the pivot adjustment knobs The heart rate displays 0 during training 1 The heart rate m...

Page 23: ...n a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to ra...

Page 24: ...th 0 5kV 1 kV line s to lines 0 5kV 1 kV 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations...

Page 25: ...ronment guidance Conducted RF IEC 61000 4 6 3 Vrms 150 kHz 80 MHz 6 V rms in ISM bands between 150 kHz and 80 MHz 80 AM at 1 kHz 3 Vrms 150 kHz 80 MHz 6 V rms in ISM bands between 150 kHz and 80 MHz 8...

Page 26: ...LTE Band 5 Pulse Modulation b 18 Hz 2 0 3 28 1720 1845 1970 1700 to 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse Modulation b 217 Hz 2 0 3 28 2450 2400 to 2570 Bluetooth WLAN 80...

Page 27: ......

Page 28: ......

Page 29: ......

Page 30: ...3428 20 09 C 2020 Compass Health Brands Corp Manufactured for Compass Health Brands Corp Toll Free 1 888 549 4945 6753 Engle Road Middleburg Heights OH 44130 richmarweb com...