Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment-guidance
(for home and professional healthcare
environment)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.