Adjustments and Calibration
7
3.3
Calibration Procedure
Refer to the indicator manual to determine correct calibration procedures.
It is recommended the scale be exercised before calibration to ensure everything is seated.
1. Load the scale to near capacity two or three times.
2. With no load on the scale, place the indicator in calibration mode and perform a zero calibration.
3. Place test weights on the platform equal to 70% - 80% of the scale’s capacity. If several weights are used,
distribute them evenly around the platform.
4. Perform a span calibration.
5. Remove the test weights and check the zero reading.
6. Repeat the calibration process, if necessary.