Maintenance
© Rice Lake Weighing Systems
●
All Rights Reserved
15
5.0
Maintenance
5.1
Troubleshooting and Testing
Refer to the following to check and correct any failure before contacting service personnel.
Table 5-1. Troubleshooting Table for the Rice Lake Scale Line
Scale does not turn on
Dead battery
Connect the scale to a power source.
Faulty electrical outlet
Use a different electrical outlet.
Bad power supply
Replace adaptor.
Questionable weight or the
scale does not zero
External object is interfering with the
scale.
Remove the interfering object from the scale.
Display did not show 0.0 before weigh-
ing.
Help the patient off the scale, zero the scale and begin the weighing process again.
Scale is not placed on a level floor.
Ensure the scale is level and begin the weighing process again.
Scale is out of calibration.
Check the weight with a known weight value.
The display shows E messages as detailed below.
E06
Identifier
—
ADC
AD too high
E07
AD too low
E10
Overload
Scale has been overloaded. Remove load from scale.
E4L
BAT
Battery low but still usable
—
One bar left on the indicator display.
E4U
Battery low and unstable
—
no bars left on the indicator display.
E11
CAL
Calibration Error
—
recalibrate the scale again.
Err 2
Low saturation state (low A/D)
The load cell is not connected properly. Check the cables and mechanical connec-
tions. If the problem persists, replace the set of load cells.
Err 3
High saturation state (high A/D)
See Err 2
Err 6
Unstable weight. Cannot calibrate.
Check the load cells’ mechanical surroundings and see that nothing touches them
and that the cables are properly welded.
SAT
Damaged load cell cable
Replace load cell cable.
5.2
Maintenance
The following section provides instructions for maintaining and cleaning the Rice Lake line of scales. Maintenance operations
other than those described in this section should be performed by qualified service personnel.
5.2.1
Basic Maintenance
Before the first use of the scale and after periods of non-use, check the scale for proper operation and function. If the scale
does not operate correctly, contact qualified service personnel.
Go through the following steps for basic maintenance.
1. Check the overall appearance of the entire scale for any obvious signs of damage.
2. Inspect the condition of the AC adapter for cord cracking or fraying or for broken or bent prongs.
5.2.2
Cleaning
Proper care and cleaning is essential to ensure a long life of accurate and effective operation. Before beginning the cleaning
process, disconnect the scale from the AC power source.
1. Clean all external surfaces with a clean, damp cloth or tissue. Mild soap and water solution may be used. Dry with a
clean soft cloth.
2. Do not immerse the scale in cleaning or other liquid solutions.
3. Do not use Isopropyl alcohol or other solutions to clean the display surface.
Symptom
Possible Cause
Corrective Action