Mechanical Physician Scale Installation and Operation Instructions - Troubleshooting

9

4.0

Troubleshooting

The accuracy of the Mechanical Physician Scale, when calibrated at the factory, falls within plus or minus 1/4 
pound accuracy. For the most accurate weight, always use the scale on a hard, level surface and stand in the 
center of the scale platform with the weight distributed as evenly as possible. If an error occurs or seems 
excessive, check the following:

5.0

Specifications

Problem

Possible Fix

Zero balance of beam

The weigh beam must be balanced so the pointer comes to a rest in the 
center of the eye loop (shown in Figure 3-1) when both poise weights are set 
at zero (see Figure 1-1 for poise weight location). Adjust the balance by 
turning the zero adjusting screw at the left end of the weigh beam (see 
Figure 3-1). 

Beam does not move freely

Check if the beam pointer is touching the side of the eye loop in its range of 
travel.
Visually ensure that the linkage is centered and properly aligned. Occasionally 
during shipping, the alignment will become skewed. The linkage must be free 
floating in order for the scale to weigh properly. The linkage can be seen by 
looking up underneath the top weigh beam if the scale is slightly tipped to one 
side.

Platform rocks excessively

When you push down on any corner of the platform, you should not feel any 
significant rocking. Ensure scale is setting on a level surface.

Beam does not move at all during a weighment

This can be caused by the poise weights being set to a higher weight than the 
person’s actual weight. Reset the weights to a lower weight. Also check to 
see that the steelyard rod is properly connected and aligned (see Figure 2-7).

Scale is out of calibration

Recalibrate the scale by placing a known weight on the scale and moving the 
zero adjusting screw to bring the scale back to zero. 

Table 4-1. Troubleshooting Guide

RL-MPS

RL-MPS-10

RL-MPS-20

Maximum Capacity

440 lb (200 kg)

450 lb (204 kg)

440 lb (200 kg)

Graduation

0.25 lb (0.1 kg)

Height Rod Range

24-84" (60-212 cm)

N/A

Min. Value of Height per Division

0.125" (0.1 cm)

N/A

Platform Size

14.75" x 10.75" (374.65 x 273.05 mm) 

Table 5-1. Model Specifications

Summary of Contents for RL-MPS

Page 1: ...102696 Rice Lake Mechanical Physician Scale Model RL MPS RL MPS 10 RS MPS 20 Operation Instructions...

Page 2: ......

Page 3: ...ent 7 4 0 Troubleshooting 8 5 0 Specifications 8 Technical training seminars are available through Rice Lake Weighing Systems Course descriptions and dates can be viewed at www ricelake com or obtaine...

Page 4: ...chanical Physician Installation and Operation Instructions Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webi...

Page 5: ...uminum height rod and rear wheels These installation and operating instructions will guide the user through basic instructions on pillar assembly wheel installation pillar height rod installation and...

Page 6: ...llar Installation Prior to assembly you ll need the following tools to finish assembling the Mechanical Physician Scale Phillips head screwdriver Wrench included The pillar comes separate from the sca...

Page 7: ...sembly ensuring that the RLWS logo located on the weight beam faces the scale base Figure 2 3 Attach the Pillar to the Scale Base 4 Using a phillips head screwdriver screw in the eight bolts and washe...

Page 8: ...in Figure 2 5 3 Insert the wrench included to the small hole in front of the steelyard rod and pull the hook of the steelyard rod with the hook of the wrench to connect it to the scale base as shown i...

Page 9: ...tructions 5 the top weigh beam if the scale is slightly tipped to one side Figure 2 7 View Looking up at Linkage up Underneath the Top Weigh Beam 5 Place the scale gently back on the floor Force lever...

Page 10: ...slotted holes of the height rod into the two bolts on the front side of the pillar as shown in Figure 2 8 One is located at the top of the pillar and the other towards the bottom of the pillar Figure...

Page 11: ...he wheels upside down can create scale errors 1 Align the angle iron of the wheel base to the scale platform as shown in Figure 2 9 Figure 2 9 Attach the Scale Wheels onto the Scale Base 2 Use the scr...

Page 12: ...olding the scale pointer with your finger until it touches the bottom of the eye loop area Release the scale pointer and let it rise freely up and down between the eye loop area until the scale pointe...

Page 13: ...and properly aligned Occasionally during shipping the alignment will become skewed The linkage must be free floating in order for the scale to weigh properly The linkage can be seen by looking up und...

Page 14: ...0 p m Monday through Friday and 8 a m to 12 noon on Saturday CST Telephone Sales Technical Support 800 472 6703 Canadian and Mexican Customers 800 321 6703 International 715 234 9171 Fax Fax Number 71...

Page 15: ...quipment has not been modified altered or changed by any person other than RLWS or its duly authorized repair agents RLWS will have a reasonable time to repair or replace the defective equipment Buyer...

Page 16: ...12 Mechanical Physician Scale Installation and Operation Instructions...

Page 17: ......

Page 18: ...PN 102696 11 12...

Reviews: