Tracer AV Installation Manual
6.3
Continuous Output (Stream) Format
Figure 6-1 shows the continuous output format sent from the
Tracer AV
Serial port to the Remote Display.
Note: In Setup mode, the text display message follows the following format: <100DM><Time M/S><Text!> or
<100DMText!>.
Figure 6-1. Continuous Output Data Format
<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>
ASCII 02
(decimal)
Weight data: 7 digits, right-justified, with
decimal point, l eading zer o s upr e ssion.
Overload = ^^^^^^^
Underrange = ] ] ] ] ] ] ]
Display overflow = OVERFL
LB = pounds
KG = kilogram
TN = ton
GM = grams
OZ = ounces
<space> = none
G = Gross
N = Net
Status:
<
space
> = valid
I = Invalid
M = Motion
O = Over/under range
Polarity:
<
space
> = Positive
<–> = Negative
<^> = Overload
<]> = Underrange
<CR> <LF>
or <CR>
ASCII 13, 10
(decimal)
Z = Center of zero