Fieldbus Options Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
75
10.3 EtherNet/IP Configuration
NOTE: The 682 must be configured for Fieldbus and EtherNet/IP:
• Setup → Communications → Serial → RS-485 or Serial Option Card Port set to Fieldbus
• Setup → Communications → Fieldbus → Network Protocol set to EtherNet/IP
10.3.1 EDS File Setup
This procedure walks through setting up the EtherNet/IP with an EDS file, preferred method. Use RXLogix to import EDS files.
1. Select the
Module Type Vendor Filter
.
2. Select
Hilsher GmbH
.
3. Select the catalog number
NIC 52-RE/EIS
and click Change.
4. Select
Exclusive Owner
from the drop down list and click
OK
.
10.3.2 Generic Module Setup
Use the following settings to setup a generic module.
• Comm Format:
Data - DINT
•
Input:
Assembly Instance: 101 | Size: 33
•
Output:
Assembly Instance: 100 | Size: 32
•
Configuration:
Assembly Instance: 8 | Size: 0
10.4 PROFINET Configuration
NOTE: The 682 must be configured for Fieldbus and PROFINET:
• Setup → Communications → Serial → RS-485 or Serial Option Card Port set to Fieldbus
• Setup → Communications → Fieldbus → Network Protocol set to PROFINET
Download GSDML files from the Rice Lake Weighing Systems website and configure PROFINET based on
Figure 10-2. PROFINET Device Overview
NOTE: The following sections apply for both EtherNet/IP and PROFINET.