MSI-4300 Port-A-Weigh Plus
10
Visit our website
4.0
Configuration
This section provides an overview of the MSI-4300 configuration instructions.
4.1 Navigation
Figure 4-1. Navigation Keys
4.2 Calibration
There are three modes of calibration available:
• Standard calibration
• Fine calibration
• Calibration with the resistor calibration number (
RCal
)
The following section instructs the user to perform each one.
4.2.1 Enable Calibration/Setup
1. Remove the seal port on the right side of the MSI-4300 front casting.
2. Select the small button located inside the seal port.
See
for navigation keys to use this menu.
Item No.
Description
1
Power – Escape calibration/setup
2
Zero – Zero the scale during calibration;
CLEAR
during setups
3
Net/Gross – Scroll through menu items; Used in
FINE CAL
as
UP
4
Tare In/Out – Used only in
FINE CAL
as
DOWN
5
Test – Used as
ENTER
; When desired menu item displayed, press to store item
Table 4-1. Navigational Keys
Parameter
Selections
FILTER
Low, Med, High
UNITS
kg, lb, Ton, Metric Ton
STD CAL
Starts Cal procedure
FINE CAL
Minor calibration adjustments
CAL RCAL
Emergency calibration only
STANDARD
Industry, NIST, OIML, Alaska, Canada, United Kingdom, Legal-for-Trade, Australia
AUTO OFF
Enabled, Disabled
AZM
Enabled, Disable
REMOTE
Enabled, Disabled
LIGHTING
Automatic, On, Off
CONTRAST
Eight steps with numerics at “123456” and message reading “Contrast”
EXIT
Select Enter (Test) to quit setup
HI SET PT
Hi Value Disabled, Gross, Netgross Response Normal, Blanklcd, Blinklcd Exit
LO SET PT
Lo Value Input Wt Mode Disabled, Gross, Netgross Response Normal, Blanklcd, Blinklcd Exit
TOTALIZE
Manual, Auto
Table 4-2. Calibrate Setup Menu
T
0
ZERO NET/GROSS TARE
T
S
E
T
R
E
W
O
P
IN / OUT
1
2
3
4
5