Rice Lake EZ Mount Installation Manual Download Page 9

Mechanical Installation

© 

Rice Lake Weighing Systems 

 All Rights Reserved

5

Figure 2-2. Orientation

2.

Assemble the mounts by inserting either a dummy load cell or the actual 
load cell into the hole in the top plate loading bracket, then installing the 
load pin and locating pin

*

. Then pass the load cell bolts through the ends 

of the load cell and load bars and thread into the base plate. Tighten by 
hand.

  The arrow on the load cell should point in the direction of the load.

3.

Lift and block the vessel to the same height as the assembled mounts.

4.

Remove the block from one support point and slide mount into position.

5.

If the mount is being fitted under the leg of a vessel, verify that the leg’s 
center line passes through the center of the top plate (through the center 
of the load cell).

6.

Lower the corner or side of the vessel carefully onto the top plate. The 
force of a vessel weighing several tons can damage a load cell if 
dropped only a fraction of an inch.

7.

With the top plate positioned approximately level, mark holes for 
attaching the top plate to the vessel’s mounting surface. Drill holes and 
attach top plate loosely to vessel with suitable fasteners.

8.

Repeat steps 3-7 for the mounting assemblies at the remaining corners 
or sides.

9.

Verify that there is no initial misalignment between the base plate and 
top plate and that the load cell is in the center of the hole in the top plate 
loading bracket. Relocate if necessary.

10. Attach the base plates to the foundation using suitable anchors for 

concrete or by bolting or welding to a steel structure. Verify that the base 
plates are no more than ±0.5° out of level. Shim as necessary.

Note

Summary of Contents for EZ Mount

Page 1: ...EZ Mount Load Cell Mounting Kit Installation Guide PN 25710 Rev C March 10 2018...

Page 2: ...marks of their respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserv...

Page 3: ...tes can be viewed at www ricelake com training or obtained by calling 715 234 9171 and asking for the training department 1 0 Introduction 1 1 1 Safety 2 2 0 Mechanical Installation 3 2 1 General Inst...

Page 4: ...Load Cell Mounting Kit ii Visit our website www RiceLake com Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com w...

Page 5: ...n the load bearing groove of the cell This design is very effective in providing for thermal expansion contraction with little friction In the majority of applications the assemblies are self checking...

Page 6: ...te injury Indicates information about procedures that if not observed could result in damage to equipment or corruption to and loss of data General Safety Do not operate or work on this equipment unle...

Page 7: ...where necessary to achieve correct load distribution If the actual load cells are used during installation take extreme care to prevent overload damage A tank or hopper can exert huge forces when drop...

Page 8: ...assembly Carefully consider three areas which commonly cause accuracy problems Are the supporting legs adequately braced so they will not spread when the system is fully loaded Does the supporting str...

Page 9: ...e center of the top plate through the center of the load cell 6 Lower the corner or side of the vessel carefully onto the top plate The force of a vessel weighing several tons can damage a load cell i...

Page 10: ...l load distribution make final height adjustments by placing shims between the top plate loading bracket and the weighing vessel The variation in load among the cells should be no more than 20 To accu...

Page 11: ...re not the vessel 3 If conduit protection is necessary against mechanical or rodent damage to the load cell cables use flexible conduit and conduit adapters at the load cells Conduit can also provide...

Page 12: ...m 4 0 Junction Box Connections Adjustments Calibration Refer to Junction Box manual for trimming details Refer to indicator manual or Technical Information section in Rice Lake Weighing Systems Load C...

Page 13: ...ebris restricting load cell movement or debris between scale and structure 2 Check that tank vessel and mounts are plumb level and square at critical areas 3 Check all piping and conduit for connectio...

Page 14: ...n that load cell is disconnected and replaced with simulator Symptom Possible Cause No return to zero Mechanical binding or debris in seals or under load cells May have lost system calibration Non lin...

Page 15: ...14773 14795 14797 14797 26071 3 Washer 6 15177 15184 15184 15184 26070 4 Double Ended Shear Beam Load Cell 1 See Load Cell Selection Guide 5 Load Bar 2 18267 18268 18269 18269 26073 6 Base Plate 1 18...

Page 16: ...acity of 200 000 lb and 250 000 lb Item No Description Qty Replacement Part Numbers A B C D 1 Top Plate Loading Bracket 1 18376 18377 18378 29013 2 Load Cell Bolt 2 14774 14796 14798 14798 3 Washer 6...

Page 17: ...Replacement Parts Rice Lake Weighing Systems All Rights Reserved 13...

Page 18: ...EZ Mount Load Cell Mounting Kit 14 Visit our website www RiceLake com...

Page 19: ......

Page 20: ...anada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems is an ISO 9001 registered company Rice Lake Weighing Systems Speci cations subject to change wit...

Reviews: