Rice Lake CLS-M2 Troubleshooting Manual Download Page 4

4

CLS-M2 Troubleshooting Guide

Loctite is a registered trademark of Henkel Corporation, U.S.A.

VIRTUi2 displays INVALID.

Calibration switch in open position
J-box error check wiring connections
Check load cell wires
Check excitation on j-box at 4.1.

Move calibrationswitch to closed position.
Press reset in VIRTUi2.
Correct wiring.
Change j-box.

VIRTUi2 weight display is blank.

No power.

Turn on COMPOW
Check

Scale carriage is not fitting 
securely on forklift during 
installation.

May have to use electric grinder to grind 
down the centering pin on the forklift scale or 
center slot on forklift.

Discuss with local terminal manager to 
determine if scale dealer is to perform this 
chargeable service.

May have to grind or torch down the area 
where the top cleats of the forklift scale are 
mounted.

May have to grind or torch side of forklift 
carriage due to previously installed side shift 
protection method.

Loose screws on j-box

Apply Loctite

® 

242 or 243 for greasy 

enviroments, to screws and tighten.

Symptom

Possible Cause

Action