VIRTUi2 displays INVALID.
Calibration switch in open position
J-box error check wiring connections
Check load cell wires
Check excitation on j-box at 4.1.
Move calibrationswitch to closed position.
Press reset in VIRTUi2.
Correct wiring.
Change j-box.
VIRTUi2 weight display is blank.
No power.
Turn on COMPOW
Check
Scale carriage is not fitting
securely on forklift during
installation.
May have to use electric grinder to grind
down the centering pin on the forklift scale or
center slot on forklift.
Discuss with local terminal manager to
determine if scale dealer is to perform this
chargeable service.
May have to grind or torch down the area
where the top cleats of the forklift scale are
mounted.
May have to grind or torch side of forklift
carriage due to previously installed side shift
protection method.
Loose screws on j-box
Apply Loctite
®
242 or 243 for greasy
enviroments, to screws and tighten.
Symptom
Possible Cause
Action