Introduction

© 

Rice Lake Weighing Systems 

 All Rights Reserved

9

1.5.5 Preset Tare (Keyed Tare)

1. With the scale empty and display showing zero weight, press 

.

2. Display will show (000000); the focused digit will flash.
3. Edit the value using the following method; or with the 

480PLUS

, use the keypad.

• Press   or   to select the digit. 
• Press   or   to increment or decrement the value. 

4. Press 

 when the value is correct. The display will change to the Net mode 

and the 

PT

 LED lights to show the preset tare was entered.

1.5.6 Display Tare

When a stored Tare value is displayed, the Gross and Net LEDs will be off and the 

 

will be lit. To display a stored tare:

1. Press 

.

2. Press   to AUDIT.
3. Press   to 

TARE

 and press  . 

4. Press   repeatedly to return to weighing mode.

If there is no tare in the system, the value displayed will be zero and the Gross and Net 

LED will be turned off.

See 

Section 3.2 on page 22

 for more information.

TARE

TARE

0

0

MENU

Summary of Contents for 480 Legend Series

Page 1: ...PRESENTED BY JPBOWLIN COM 817 332 8116 J P Bowlin is a quality driven Calibration Company that has provided calibration repair and sales of all types of weighing and measurement products since 1931 ...

Page 2: ...480 Legend Series Digital Weight Indicator Version 1 05 Operation Manual PN 163374 Rev D October 3 2019 ...

Page 3: ...emarks of their respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserves the right to make changes to the technology features specifications and design of the equipment without notice The most current version of this publication software firmware and all other product ...

Page 4: ...s 6 1 4 2 Edit Parameter Values 7 1 4 3 Numeric Keypad Editing Procedure 480Plus Only 7 1 5 Indicator Operations 8 1 5 1 Status Lights While in Various Menus 8 1 5 2 Zero Scale 8 1 5 3 Toggle Units 8 1 5 4 Acquire Tare 8 1 5 5 Preset Tare Keyed Tare 9 1 5 6 Display Tare 9 1 5 7 Print Ticket 10 1 5 8 Toggle Gross Net Mode 10 1 5 9 View Audit Trail 10 1 5 10 Enter New Unit ID 11 1 5 11 Display Accum...

Page 5: ...eries Digital Weight Indicator ii Visit our website www RiceLake com Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webinars ...

Page 6: ...utput module provides 0 10 2 10 VDC or 0 20 4 20 mA tracking of gross or net weight values Optional digital I O card four outputs two inputs for setpoints and key functions Unit ID up to six numeric operator entered Accumulator with report and clear Time and date Audit trail tracking Supported Applications Custom Ticket Printing Gross Net Setpoint format can be customized up to 300 characters and ...

Page 7: ...uipment unless this manual has been read and all instructions are understood Contact any Rice Lake Weighing Systems dealer for replacement manuals Failure to heed may result in serious injury or death DO NOT open the indicator all procedures that require work inside the indicator enclosure are to be performed by qualified service personnel only DO NOT allow minors children or inexperienced persons...

Page 8: ...ion It is accessible by pressing the MENU key on the front panel 1 3 Front Panel Display Figure 1 1 shows the 480 LED annunciators keypad and key functions The symbols shown by the keys representing up down enter left right describe the key functions assigned in the operating modes The keys are used to navigate through menus select digits within numeric values and increment decrement values Figure...

Page 9: ...ll scale but can be configured for up to 100 of full scale Switches the weight display to an alternate unit In numeric entry mode used as a clear key Sends on demand print format out the serial port provided the conditions for standstill are met PRINT may be displayed while the unit prints Performs one of several predetermined Tare functions dependent on the mode of operation s To view a stored ta...

Page 10: ...se that can be displayed or printed Standstill LED Scale is at standstill or within the specified motion band Some operations including Zero Tare and Printing can only be done when the standstill LED is on lb kg LED Displays which unit of measure is being used lb and kg annunciators indicate the units associated with the displayed value lb pounds kg kilograms The displayed units can also be set to...

Page 11: ...ation mode 1 4 1 Navigating Through Levels Figure 1 3 Menu Navigation To select a parameter press or to scroll left or right until the desired menu group appears on the display then press to move down to the sub menu or parameter you want When moving through the menu parameters the present value appears first on the display PRINT MENU UNIT POWER TARE GROSS NET B N ZERO Move UP Increment Value Move...

Page 12: ... keypad option the method for editing numeric values relies on the numbers which are embossed on the keypad in oppose to using the arrows 1 When editing numeric values insert the required value using the numeric keypad 2 Press to save the value entered and return to the level above Press to set the currently selected digit to 0 Press to enter a decimal point When editing fractional numeric values ...

Page 13: ...urth sub menu level Figure 1 6 Status Lights 1 5 2 Zero Scale 1 In gross mode remove all weight from the scale and wait for the LED to light 2 Press The LED lights to indicate the scale is zeroed 1 5 3 Toggle Units 1 Press to toggle between primary and secondary units The current unit LED will be lit 1 5 4 Acquire Tare 1 Place container on scale and wait for the LED to light 2 Press to acquire the...

Page 14: ...the value 4 Press when the value is correct The display will change to the Net mode and the PT LED lights to show the preset tare was entered 1 5 6 Display Tare When a stored Tare value is displayed the Gross and Net LEDs will be off and the will be lit To display a stored tare 1 Press 2 Press to AUDIT 3 Press to TARE and press 4 Press repeatedly to return to weighing mode If there is no tare in t...

Page 15: ...ess is ignored 1 5 8 Toggle Gross Net Mode 1 Press to switch the display mode between gross and net If a tare value has been entered or acquired the net value is the gross weight minus the tare Gross mode Gross Brutto LED is lit Net mode Net LED is lit 1 5 9 View Audit Trail 1 Press 2 Press to AUDIT 3 Press The audit trail CALIB is displayed 4 Press then or to CNT TIME or DATE 5 Press to view sele...

Page 16: ... to weighing mode 1 5 11 Display Accumulator 1 Press 2 Press to AUDIT 3 Press until display reads ACCUM 4 Press to display VIEW 5 Press or to select desired parameter VIEW TIME DATE PRINT CLR Y For VIEW TIME or DATE press to view the value Press or to return to selected parameter To PRINT or CLEAR press then press to print or clear the accumulator Press to return to selected parameter 6 Press repe...

Page 17: ...or 12 hour format hh mm ss use the following method Press or to select hours minutes or seconds the selected value will be flashing Press or to increment or decrement the value 7 Press when the value is correct Use the same procedure to enter the date in the same format configured for the indicator 8 Press repeatedly to return to weighing mode The time and date are backed up with an internal batte...

Page 18: ...iew and edit the value To edit Value use the following method or with the 480PLUS use the keypad Press or to select the digit Press or to increment or decrement the value Press when the value is correct To edit ENABLE Press or to select ON OFF Press when the value is correct 8 Press repeatedly to return to weighing mode 1 5 14 View Firmware Version 1 Press 2 Press AUDIT is displayed 3 Press until ...

Page 19: ...ayed 4 Press CNFG is displayed 5 Press to USER 6 Press 000000 is displayed 7 To edit the password use the following method or with the 480PLUS use the keypad Press or to select the digit Press or to increment or decrement the value Press when the value is correct 8 Press to return to PASWRD 9 Press to CONFIG 10 Press to return to weighing mode When entering a user function the operator will now be...

Page 20: ...TEST VERS SERIAL ETHNET USB See Figure 2 2 See Figure 2 3 on page 16 See Figure 2 4 on page 17 See Figure 2 5 on page 18 See Figure 2 6 on page 18 See Figure 2 8 on page 19 See Figure 2 7 on page 19 TIMDAT ACCUM SERIAL SETPNT ETHNET BATCHG SETPT1 8 USER 000000 PREACT SUPVSR CONFIG FORMAT CALIBR PROGRM DIG IN ALGOUT PASWRD USBMEM MENU TEST AUTO OFF MANUAL 000000 VALUE OFF ON ENABLE OFF ON PUSHAC OF...

Page 21: ... REMOTE NONE STRUR STRLFT STRIND ECHO OFF ON SBITS 1 STOP 2 STOP TERMIN CR CR LF EOLDLY 000 BITS 7ODD 7EVEN 8NONE PRNMSG OFF ON BAUD 4800 2400 1200 19200 38400 9600 BITS 7ODD 7EVEN 8NONE SBITS 1 STOP 2 STOP TERMIN CR CR LF EOLDLY 000 ECHO OFF ON PRNMSG OFF ON TRIGER STR2 STR3 STR4 NONE PRN STR1 STR5 4800 2400 1200 19200 38400 9600 MENU CONFIG FORMAT CALIBR PROGRM DIG IN ALGOUT PASWRD USBMEM TEST ...

Page 22: ... 0 0 0 0 ON OFF 0 0 0 0 0 0 0 0 00 00 00 00 00 00 DHCP STRIND STRLFT COMAND DEMAND TRIGER TIMOUT TERMIN SFMT PORT EOLDLY ECHO CR CR LF 0 0 255 1 65535 10001 ON OFF 0 65535 0 STR 1 STR 2 STR 5 STR 4 STR 3 STRIND STRLFT COMAND DEMAND TRIGER TERMIN CR CR LF EOLDLY 0 0 255 RMPORT 1 65535 1 ECHO ON OFF TIMOUT SFMT RMOTIP 0 0 0 0 MENU CONFIG FORMAT CALIBR PROGRM DIG IN ALGOUT PASWRD USBMEM TEST ...

Page 23: ...T PASWRD USBMEM TEST TIMDAT ACCUM SERIAL SETPNT PFRMAT Display first 6 characters of format Display and edit active character and ASCII value Delete active character Scroll right in format string Scroll left in format string Increment ASCII value of active character Decrement ASCII value of active character Press to insert a space before the active character GFMT NFMT ACCFMT SPFMT MENU CONFIG FORM...

Page 24: ...re 2 8 Misc Menu VERS PFRMAT MISC FIRMW LR Legally Relevant VERSION VERSION MENU CONFIG FORMAT CALIBR PROGRM DIG IN ALGOUT PASWRD USBMEM TEST MISC POWER BKLGHT BAT OFF ON AUTO 10 SEC 30 SEC 1 MIN 5 MIN 10 MIN BAT LEVEL AUTO MANUAL MENU CONFIG FORMAT CALIBR PROGRM DIG IN ALGOUT PASWRD USBMEM TEST ...

Page 25: ...ation checksum error Recalibrate load cells EIDATA Internal RAM checksum error Call Rice Lake Weighing Systems Service at 800 472 6703 E REF A D reference error A D converter requires recalibration Call Rice Lake Weighing Systems Service ERROR Internal program error Check configuration Call Rice Lake Weighing Systems Service if unable to clear error by cycling power or if error recurs OVERFL Overf...

Page 26: ...at previous power off state due to low battery or no battery condition The printer accumulator and AUDUT functions will fail to get time and date EUCKSM Configuration checksum The checksum value of configuration has changed from that stored in memory OIMLER OIML parameter error Parameter set incorrectly for use in the OIML mode Example Primary units set for lb or oz EE ACC Accumulator error Error ...

Page 27: ... tare Zero Positive No Tare Zero Yes Tare Zero Canada Zero No 000000 Zero Yes Clear tare Clear tare Negative No No action Zero Yes Clear tare Clear tare Positive No Tare Zero Yes No action Clear tare OIML Zero No 000000 Zero Yes Clear tare Zero Clear tare Negative No No action Zero Yes Clear tare Zero Clear tare Positive No 000000 Zero Yes Tare Zero Clear Tare None Zero No 000000 Zero Yes Clear ta...

Page 28: ...SRQVLELOLW WKDW WKH SURGXFWV WR ZKLFK WKLV GHFODUDWLRQ UHIHUV WR LV LQ FRQIRUPLW ZLWK WKH IROORZLQJ VWDQGDUG V RU RWKHU UHJXODWLRQV GRFXPHQW V HXWVFK LU HUNOlUHQ XQWHU XQVHUHU DOOHLQLJHQ 9HUDQWZRUWXQJ GDVV GLH 3URGXNWH DXI GLH VLFK GLHVH UNOlUXQJ EH LHKW GHQ IROJHQGHQ 1RUPHQ XQG 5HJXOLHUXQJVEHVWLPPXQJHQ HQWVSUHFKHQ UDQFDLV 1RXV GpFODURQV VRXV QRWUH UHVSRQVDELOLWp TXH OHV SURGXLWV DX TXHOV VH UDSSR...

Page 29: ... Up to 7 mV V Analog Signal Sensitivity 0 1 V graduation minimum 0 5 V grad recommended Local Resistance 35 1140 Noise ref to input 0 5 V p p Internal Resolution 523 376 counts Display Resolution 100 000 dd Measurement Rate 37 measurements sec Input Sensitivity 38 nV per internal count System Linearity Within 0 01 of full scale Zero Stability 13 nV C Span Stability 13 ppm C Calibration Method Soft...

Page 30: ...active high Digital Outputs 4 or 8 dry contact relays Up to 30VDC at 2A current Serial Communications Port 1 Full duplex RS 232 Port 2 Full duplex RS232 or output only Active 20mA current loop Both Ports 1200 to 38400 bps 7 or 8 data bits even odd or no parity 1 or 2 stop bits Operator Interface Display 6 digit LED display 7 segment 0 8 in 20 mm digits LED annunciators Gross net center of zero sta...

Page 31: ...TC8322 Accuracy Class III nmax 10 000 Measurement Canada Approval AM 5892 Accuracy Class III IIIHD nmax 10 000 UL File Nubmer 151461 FCC The 480 complies with Part 15 of the FCC Rules Operation is subject to the following conditions This device may not cause harmful interference This device must accept any interference received including interference that may cause undesired operation Warranty 2 y...

Page 32: ......

Page 33: ...anada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems is an ISO 9001 registered company Rice Lake Weighing Systems Specifications subject to change without notice www ricelake com PN 163374 Rev D October 3 2019 ...