Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
25
4.3.4 Setup – Communication Menu
Figure 4-8. Setup – Communication Menu
wlin
Linear Calibration – A linear or multi-point calibration is performed by entering up to four additional calibration
points;
WLIN V# – Sets the test weight value for linear calibration point
WLIN C# – Executes the linear calibration process for the point; generates the raw count value (F) for
the test weight value (V)
WLIN F# – Displays the raw count value at the linear point weight; A linear calibration (WLIN C#)
generates this raw count value; Manually changing this count value changes the linear point weight and
negates the linear calibration for the point
rezero
Rezero – Removes an offset value from the zero and span calibrations
LSTZRO
Last Zero – Takes the last pushbutton zero in the system (from weigh mode) and uses it as the new zero
reference point, after which a new span calibration must be performed; This calibration cannot be performed
when calibrating a scale for the first time
TMPZRO
Temporary Zero – Temporarily zeros the displayed weight of a non-empty scale, after a span calibration was
performed; The difference between the temporary zero and the previously calibrated zero value is used as an
offset
Menu
Description
serial
Serial Port – Supports RS-232 serial communications; see
USB
USB Port – Supports a USB Virtual PC Communication Port; see
Table 4-6. Setup – Communication Menu Descriptions
Parameter
Description
Table 4-5. Setup – Calibration Menu Description
Setup
Comm
SERIAL
usb