2.4 Cable Connections
The 380 provides four cord grips at the bottom of the enclosure for cabling into the indicator. One of the cord grips is used for
the load cell and the other three are used to accommodate the serial connection and other future connections. Cable plugs
included in the parts kit and must be installed in open cord grips to prevent moisture from entering the enclosure. Depending on
the application, install cables as required. The recommended cable strip length is 0.25'' (7 mm) for all 380 connectors. See
for the recommended assignments for the 380 cord grips.
Only connect unit to equipment certified to IEC 60950, IEC 62368, IEC 61010 or similar.
Figure 2-5. Recommended Cord Grip Assignments
Do not have open/bare wires outside of the enclosure. Make sure no stripped portion of cable is on the outside
of the cord grips.
Properly seal cord grips to prevent moisture damage inside of the enclosure. Cable plugs must be installed in
unused cord grips. Cord grip dome nuts, around a cable or a plug, must be torqued to 22 in-lb. The cord grip nut
against the enclosure must be torqued to 33 in-lb.
WARNING
Batteries
Cord grips for
communication cables
Power Jack
Cord grip for
load cell cable
Breather Vent
IMPORTANT