3.7 Troubleshooting

Refer to the following table to check and correct any failure before contacting service personnel.

Symptom

Possible Cause

Corrective Action

Scale does not turn on

Dead batteries

Replace batteries or connect to AC power

Faulty electrical outlet

Use a different electrical outlet

Bad power supply

Replace adapter

Questionable weight or the scale does not 

zero

External object is interfering with the scale Remove the interfering object from the scale
Display did not show 0.0 before weighing Help the patient off the scale, zero the scale and begin the 

weighing process again

Scale is not placed on a level floor

Ensure scale is level and begin the weighing process again

Scale is out of calibration

Check the weight with a certified calibration weight 

Scale base is touching floor during a 

weighment

Adjust height of feet so fingers can slide between the base 

of scale and the floor all the way around the platform

The display shows a 

STOP

 message

The load on the scale exceeds the 

capacity of the scale

Remove the excess weight and use the scale according to 

manufacture specifications

The display shows 

LO Bat

 message

The battery is low

Replace batteries

The display shows E and Err messages as detailed below
E06

Identifier - ADC

AD too high

E07

AD too low

E10

Overload

Scale has been overloaded. Remove load from scale

E4L

BAT

Battery low, but still usable- one bar left on indicator display

E4U

Battery low and unstable - no bars left on indicator display

E11

CAL

Calibration Error - recalibrate scale

Err 1

Load cell cable may be plugged into wrong 

connection port

Ensure cable is connected to the load cell connection port.

Note: 

Load cell connection point is located underneath the 

curved plastic cover of the indicator. Remove four back 

retaining screws, remove curved back cover to access load 

cell connection point.

Err 2

Low saturation state (low A/D)

The load cell is not connected properly; Check the cables 

and mechanical connections; if the problem persists, replace 

the set of load cells

Err 3

High saturation state (high A/D)

See Err 2

Err 6

Unstable weight; Cannot calibrate

Check the load cell mechanical surroundings and ensure 

nothing is contacting the load cell and that the cables are 

properly welded

Err 7

Scale isn’t moving

Make sure feet are installed on the scale. Turn the feet all 

the way in and then back them out three full turns, then level 

the scale

SAT

Damaged load cell cable

Replace load cell cable

Load cell cable may be plugged into wrong 

connection port

Ensure cable is connected to the load cell connection port.

Note: 

Load cell connection point is located underneath the 

curved plastic cover of the indicator. Remove four back 

retaining screws, remove curved back cover to access load 

cell connection point.

Table 3-2.   Troubleshooting Table

Summary of Contents for 350-10 Series

Page 1: ...350 10 Series Digital Wheelchair Scales Software Version 11525 Operation Manual Model 350 10 2 Model 350 10 3 Model 350 10 7 Model 350 10 8 PN 151640 Rev E August 17 2022...

Page 2: ...respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserves the right to...

Page 3: ...ing or obtained by calling 715 234 9171 and asking for the training department Revision History This section tracks and describes the current and previous manual revisions for awareness of major updat...

Page 4: ...mbly 3 2 1 Unpacking 3 2 1 1 Repacking Scale 3 2 2 Setting up Scale 3 2 2 1 Insert Batteries 5 2 2 2 Power Connection 5 3 0 Operation 6 3 1 Key Descriptions 6 3 2 Weighing 7 3 3 Hold Release Function...

Page 5: ...ring and recalibration All that is required is to open the packaging and unfold the scale Manuals resources and warranties are available from Rice Lake Weighing Systems website at www ricelake com hea...

Page 6: ...sport the scale while someone is on the scale Do not allow minors children or inexperienced persons to operate this scale Do not use in the presence of flammable materials Do not use this product if a...

Page 7: ...y packed with sufficient packing materials Whenever possible use the original carton and packing materials when shipping the scale back Damage caused by improper packaging is not covered by the warran...

Page 8: ...floor Not having clearance around the scale base creates inaccurate weighments Figure 2 2 Clearance Between Scale Base and Floor 6 Apply light pressure to all corners of the scale base to ensure there...

Page 9: ...e used for an extended period of time If the LO BAT indicator activates for accurate weighing replace the batteries or connect the scale to an AC power source as soon as possible 2 2 2 Power Connectio...

Page 10: ...BMI Pressing the BMI key enables access to the BMI Body Mass Index mode defaults when scale is turned on The patient is gets on the scale weight stabilizes and press the BMI key The display then asks...

Page 11: ...y the patient 1 Press to turn on the scale 0 0 appears on the display along with the ZERO annunciator 2 Place additional item s on the scale 3 Press until the display returns to 0 0 and NET annunciato...

Page 12: ...to adjust the height value 5 Press 6 The BMI value and BMI annunciator are shown on the display Press to return to the Weighing mode and the BMI function will be turned off 3 6 2 KG Mode 1 Ensure that...

Page 13: ...high E07 AD too low E10 Overload Scale has been overloaded Remove load from scale E4L BAT Battery low but still usable one bar left on indicator display E4U Battery low and unstable no bars left on in...

Page 14: ...press and hold for at least three seconds or until the scale emits two quick beeps If the scale does not beep after five seconds release as the weight was either in motion or out of range If displayin...

Page 15: ...the use of the USB driver Driver protections may need to be temporarily disabled on Windows 10 or later computers to allow for the installation of the USB driver 1 Connect the scale s indicator to a P...

Page 16: ...ice personnel Go through the following steps for basic maintenance Check the overall appearance of the entire scale for any obvious signs of damage Inspect the condition of the AC power adapter cord f...

Page 17: ...figured Data Communications RS 232 with RJ 45 jack USB Connection Selectable baud rate default 9600 8 bits No parity 1 stop bit No handshaking Environmental Operating Temperature 50 F to 95 F 10 C to...

Page 18: ...350 10 Series Digital Wheelchair Scales 14 Visit our website www RiceLake com...

Page 19: ......

Page 20: ...54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems Specifications subject to change without notice www ricelake com PN...

Reviews: