26
changed the algorithm resets to the default which should be changed to the required
algorithm before taking a measurement.
The calibrate button functions is described in section 8.6.1.
The default saving behaviour depends on the algorithm selected but usually includes
the unfiltered altitude map, any filtered map (as determined by the algorithm selection),
curvature maps and slope maps. Full saving can be enabled by pressing the
“Full
sa
ving” button
. This will include the images taken that are used for the computation
process and data that is created and used in the computation process. Saving all
measurement images significantly increases the stored data size in the TAMS memory
and reduces the number of measurements that can be stored, also the measurement
time increases by several seconds, transfer time to the SD card is also increased.
The reset memory button removes all saved measurements so desired results should
be exported first.
The O_ROUGH algorithm has additional settings that are presented when the
algorithm is selected. The type of filter can be changed by pressing the button between
type and the desired band.
Filter type
User can choose the desired filter type for the first band between High Pass Filter (HPF),
Low Pass Filter (LPF) or Band Pass Filter (BPF). See section 10.1 to know more about O-
Rough algorithm.
Low limit
This is where user can set the cut off value of the first band in case of HPF or LPF selected.
In case of BPF, this value is the low cut-off limit of the first band. This value is given in
millimeters with a resolution of 0.1mm.
High limit
User can set the high cut-off limit of the first band in case of BPF. This value is not used in
case of HPF and LPF. This value is given in millimeters with a resolution of 0.1mm.
The limits can be adjusted by 1 by pressing the double arrow and by 0.1 by pressing
the single arrow.