DISTILLED WATER CONNECTION
The system automatically siphons distilled water from its Distilled Water Reservoir located in the lower cabinet. A
1/8” diameter flexible
tube is pre-installed into the tank to siphon water from the tank bottom. It passes through
the tank’s cover screw, through a fitting
in the lower cabinet, connects through a fitting on the upper cabinet,
connecting finally to the presaturator fill pump. As these tubes and connections are pre-installed at the factory,
there are no other additional connections to make for distilled water.
3.2.5
SYSTEM DISCHARGE CONNECTION
There is a ½” pipe connection on the rear corner of the cabinet for discharge of air, water vapor, and
condensate
removed via the air drier.
If your pressurized air source is dry, and you generally operate the
humidity generator at a temperature no warmer than 25 °C, you will
likely never see any water coming from the discharge connection.
However, if you utilize the internal air compressor, or operate the
generator at chamber temperatures above 25 °C, condensation may
occur at this discharge connection. The higher the chamber temperature
and humidity, the more condensation will be present.
You may, if desired, connect a ½” NPT fitting and tube to direct this
discharge away from the system to a small drip pan or a floor drain. A
drip pan volume of 0.5 to 1 liter is generally sufficient.
3.3
INITIAL PREPARATION
3.3.1
COMPUTER CONNECTION
The system is supplied with its own built-in computer and monitor on the right-side front panel door. Enable the
computer, monitor, and internal control system by pressing the front panel power switch located below the
monitor in the upper right cabinet panel. The switch should illuminate, the monitor should activate, and the
operating system (Windows 10) should boot. Follow section 4.1 through 4.2.1 to launch the RHS Control program
and connect to the CGS-240 Humidity Generator.
Interaction with the system is via touch screen or physical mouse/keyboard. To utilize the physical
mouse/keyboard, plug them into the USB connectors located at the upper right rear of the system.
Additional connectors (HDMI and RS-232) are also available at the same location. The HDMI connector is for
attachment of an additional external monitor if desired. The RS-232 connector allows connection of an external
computer system if desired. An external computer system could be utilized to operate the generator in the event
of needed firmware updates, secondary system monitoring, or alternate data collection. This connection may also
Figure 3.10 Discharge Connection
Summary of Contents for CGS-240
Page 1: ...CGS 240 RH Systems LLC 1225 W Houston Ave Gilbert AZ 85233 Manual...
Page 98: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 98 of 148 Figure 8 2...
Page 99: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 99 of 148 Figure 8 3...
Page 100: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 100 of 148 Figure 8 4...
Page 101: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 101 of 148 Figure 8 5 Figure 8 6...
Page 102: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 102 of 148 Figure 8 7 Figure 8 8...
Page 103: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 103 of 148 Figure 8 9...
Page 104: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 104 of 148 Figure 8 10...
Page 105: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 105 of 148 Figure 8 11 Figure 8 12...
Page 106: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 106 of 148 Figure 8 13 Figure 8 14...
Page 107: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 107 of 148 Figure 8 15...
Page 108: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 108 of 148 Figure 8 16...
Page 109: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 109 of 148 Figure 8 17...
Page 110: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 110 of 148 Figure 8 18...
Page 111: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 111 of 148 Figure 8 19...
Page 112: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 112 of 148 Figure 8 20...
Page 113: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 113 of 148 Figure 8 21...
Page 114: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 114 of 148 Figure 8 22...
Page 115: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 115 of 148 Figure 8 23...
Page 117: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 117 of 148 Figure 8 25...
Page 118: ...Illustrated Parts Breakdown CGS 240 Manual 6 15 2021 Page 118 of 148 Figure 8 26...
Page 120: ...Drawings CGS 240 Manual 6 15 2021 Page 120 of 148...
Page 121: ...Drawings CGS 240 Manual 6 15 2021 Page 121 of 148 Figure 9 3 Refrigeration System...
Page 122: ...Drawings CGS 240 Manual 6 15 2021 Page 122 of 148 Figure 9 4 Pneumatic System...
Page 123: ...Drawings CGS 240 Manual 6 15 2021 Page 123 of 148 Figure 9 5 Fluid System...
Page 124: ...Drawings CGS 240 Manual 6 15 2021 Page 124 of 148 Figure 9 6 AC Distribution...
Page 125: ...Drawings CGS 240 Manual 6 15 2021 Page 125 of 148 Figure 9 7 AD0 3 LVL4...
Page 126: ...Drawings CGS 240 Manual 6 15 2021 Page 126 of 148 Figure 9 8 Power Distribution...
Page 127: ...Drawings CGS 240 Manual 6 15 2021 Page 127 of 148 Figure 9 9 RLY 1 14...
Page 128: ...Drawings CGS 240 Manual 6 15 2021 Page 128 of 148 Figure 9 10 PT 12 Temp Board...
Page 129: ...Drawings CGS 240 Manual 6 15 2021 Page 129 of 148 Figure 9 11 UART 0 3...
Page 130: ...Drawings CGS 240 Manual 6 15 2021 Page 130 of 148 Figure 9 12 Access to Presat Drain Fitting...
Page 131: ...Drawings CGS 240 Manual 6 15 2021 Page 131 of 148 Figure 9 13 Access to Sat Drain Fittings...
Page 132: ...Drawings CGS 240 Manual 6 15 2021 Page 132 of 148 Figure 9 14 Temperature Probes...
Page 133: ...Drawings CGS 240 Manual 6 15 2021 Page 133 of 148 Figure 9 15 Pressure Transducers Manifold...