Defective power cells

In case you think the power cell is defective comply with 

the following steps:

Place the hearing instrument into the charger. If you 

get a red LED then the power cell is defective.

Perform visual inspection of the hearing instrument. 

The test is considered failed if it has any of below 

criteria:

■ 

There is physical sign of deformation caused by 

bloated lithium power cell inside.

■ 

There is visible leakage from the lithium power cell.

■ 

Product is generating heat on its own or hot to the 

touch.

If any of the test steps failed, consider the power cell 

defective. Due to safety concerns the hearing instrument 

cannot be shipped anymore. Please record the serial 

number of the hearing instrument, take a photo of the 

damages, and consult with Sivantos Customer Service 

team before disposal of the product according to local 

regulations.

Summary of Contents for M-Core SR

Page 1: ......

Page 2: ...Activating 4 Charging 5 Connecting 5 Noting the serial number 5 Troubleshooting 6 Exchanging the receiver wax guard 6 Important safety information 7 General warnings 7 Defective power cells 8 Further...

Page 3: ...ls It provides important information on hearing instruments with permanently built in power cells lithium ion batteries CAUTION It is important to read this guide the user guide and the safety manual...

Page 4: ...clicking it in place X Push the receiver locking bar into the housing Activating Before first use or first fitting you have to activate the hearing instruments Place the hearing instruments in the ch...

Page 5: ...ruments before delivery to the client A full charging cycle takes approximately 4 hours Follow the operating conditions in both the hearing instrument s and the charger s user guide Connecting Use Noa...

Page 6: ...exchange the receiver wax guard Exchanging the receiver wax guard X Remove the ear piece X Screw the threaded end of the wax guard changing tool into the filter of the external receiver X Remove the u...

Page 7: ...nstruments NOTICE X Do not exchange power cells This may only be performed by Service Center only General warnings WARNING A power cell lithium ion rechargeable battery is permanently built into the h...

Page 8: ...re is physical sign of deformation caused by bloated lithium power cell inside There is visible leakage from the lithium power cell Product is generating heat on its own or hot to the touch If any of...

Page 9: ...s after production date The production date is printed on the hearing instrument s label on the packaging Once a hearing instrument is activated in order for it to remain in good condition it needs to...

Page 10: ...to iPhone iPad or iPod respectively and has been certified by the developer to meet Apple performance standards Apple is not responsible for the operation of this device or its compliance with safety...

Page 11: ...11 Risk mitigations for internal use only Defective power cells 8 MI301 Follow safety instructions for Li ion 7 Not exchange power cells 7...

Page 12: ...No 04229 99T Order Item No Master Rev01 05 2020 2020 Sivantos GmbH All rights reserved www rexton com 0123 Legal Manufacturer Sivantos GmbH Henri Dunant Strasse 100 91058 Erlangen Germany Phone 49 91...

Reviews: