background image

meaSUriNg methoDS

in the mouth (oral)

Position the thermometer in one of the two pockets under the tongue, to the left or right of the root of the tongue. The 

measuring sensor must be in good contact with the tissue. Close your mouth and breath evenly through the nose to 

prevent the measurement from being influenced by inhaled/exhaled air. 

Approx. measuring time: 1 minute.

in the anus (rectal)

This is the most reliable measuring method, and is especially appropriate for infants and small children. Carefully insert 

the tip of the thermometer 2 to 3 cm into the anal aperture. 

Approx. measuring time: 1 minute.

in the armpit (axillary)

From a medical viewpoint, this method will always provide inaccurate readings, and should not be used if precise 

measurements are required.

Approx. measuring time: 5 to 10 minutes.

CleaNiNg aND DiSiNFeCtiNg

The thermometer is 100% waterproof and may be immersed in water or disinfectant solution, 

then wipe it off with a moist towel. Please use one of the following disinfectants for cleaning, following the manu-

facturer’s instructions in each case.

Name of Disinfectant

Dilution with Water

max. Duration method

Isopropyl alcohol 70%

95.0%

Immerse for 24 hours

Note:

 

any other cleaning solution, dilution or cleaning method can cause a malfunction of the thermometer. 

the thermometer must never be boiled.

BatterY DiSPoSal

Please break the end cap from the main body with a pair of pliers. Remove the battery and the PC board from the plastic 

housing and dispose of separately. Battery cannot be replaced.

teChNiCal SPeCiFiCatioNS

type

Maximum thermometer

measurement range

32.0°C to 43.9°C / 89.6°F to 111°F

Temp. < 32.0°C / 89.6°F; display «

l

» for low (too low)

Temp. > 43.9°C / 111°F; display «

h

» for low (too high)

measurement accuracy

±0.1°C / 0.2°F between 34°C / 93.2°F and 42°C / 107.6°F at an 

ambient temperature of 18°C / 64.4°F to 28°C / 82.4°F

Self-test

Automatic internal check at a test value of 37.0°C / 98.6°F; if there is a 

deviation of >0.1°C / 0.2°F, «

err

» (error) is displayed.

Display

Liquid crystal display (LCD) with three digits

Smallest unit of display: 0.1°C / 0.1°F

memory

For storing the last measured value

operating environment

10°C to 40°C / 50°F to 104°F

Storage temperature

-25°C to 60°C / -13°F to 140°F

Storage humidity

15% to 95% (noncondensing)

Battery

IEC type SR41 Silver oxide high drain

Battery lifetime

Approx.4500 measurements

reference to standards

EN12470-3, clinical thermometers

ASTM E 1112, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11

liFetime WarraNtY

This product is warrantied to be free from defects in material or workmanship for the life of the original owner.  It will be 

repaired or replaced without charge upon receipt of the unit prepaid with $5.00 to cover handling, packaging and return 

postage to the address below.  Please include your full name, address and daytime phone number or email address.

Type BF applied part

Batteries and electronic devices must be disposed of in 

accordance with the locally applicable regulations, not with 

domestic waste.
Follow Instructions for Use. This document provides important 

product operation and safety information. Please read this 

document thoroughly before using the device and keep for 

future reference.

IP22: Protected against solid foreign objects of 12.5 mm diameter and 

greater. Protected against vertically falling water drops when the device is 

tilted up to 15°.