DUROMETER OPERATING INSTRUCTIONS 

To operate the durometer, press the foot of the gauge firmly against the specimen, but not so firmly as to 
imbed the foot into the surface. This will yield the durometer reading.  

PRIMARY BUTTON FUNCTIONS 

ON ZERO

                        Powers durometer on. Zero / clears reading. 

a.

 

To Power Durometer On: Press and Release 

ON ZERO

 

b.

 

To Clear / Zero Display: Press and Release

 ON ZERO 

when LCD is on  

(Zero in release on button) 

OFF MODE

                        Powers off.  

a.

 

To Power Durometer Off: Press and Release 

OFF MODE

 

(Value and settings are not lost when durometer is off) 

HOLD

 

 

Toggles HOLD on and off.  

a.

 

To Toggle HOLD On and Off: Press and Release 

HOLD

 

2ND

 

 

Enables 2

ND

 Functions (DATA) and 3RD Functions (LOCK) 

TOL

 

 

Toggles TOL (Tolerances) on and off. 

a.

 

To Toggle TOL On and Off: Press and Release 

TOL

 

If no tolerances are programmed, a large tol will be displayed and both the 

LOW and HI icons will flash. 

b.

 

To Program LOW and HI Tolerances: Toggle 

TOL 

On (Step A). Press and Hold 

TOL 

until HI is flashing and release 

TOL. Use 

move

 (2

ND

), 

change

 (TOL

), apply

 

(HOLD)

 to program HI tolerance

Move

 (2

ND

scrolls flashing icon left to right.

 

Change

 (TOL) 

toggles value of flashing icon. 

Apply 

(HOLD)

 stores value HI or 

LOW

.

 

c.

 

When the value is out of tolerance the numbers will flash and the LOW or HI 

will flash. If the LEDs are turned on, the green or red LEDs will flash for 2 

seconds when in tolerance and 4 seconds when out of tolerance. 

d.

 

To Disable / Enable LEDs for TOL: USE the Following 3 Button Sequence 

1.

 

Press and release 

2

ND

 

2.

 

Press and Release 

ON ZERO

 

3.

 

Press and Hold 

INmm

 Until Display Changes 

(All On, Off, DATA or TOL will be displayed) 

Use 

change

 (TOL)

 to scroll All On (LEDs will flash when using the TOL 

and DATA), Off (turned off for both TOL and DATA), TOL (turned on for 

tolerance only), DATA (turned on for DATA only). Use 

apply

 (HOLD)

 to 

store desired setting. 

Note: Factor reset: Press and Hold 

ON ZERO

 and 

OFF

 simultaneously 

until RESET appears on LCD. 

 

 

 

2ND BUTTON FUNCTIONS 

DATA

 

 

Send data through cable, wireless radio module, or stored readings in durometer. 

a.

 

To Send Data or Store Readings 

1.

 

Press and Release 

2

ND

 

2.

 

Press and Release

 DATA (HOLD)

 

Cable: If cable is connected, the data will be sent through the data 

output cable. For 1 second DATA icon will appear.  

Note: USB Data Cable 

G13-0055

 - Sold Separately 

Wireless: If wireless output is enabled, the durometer reading will be 

sent by wireless output. For 1 second DATA icon will appear. 

Stored Reading: With no cable connection and wireless not enabled, 

readings will be stored in durometer. For 1 second DATA icon will 

appear. 

Note: A computer can also request readings and you can setup the 
durometer to automatically send readings over wireless. 

b.

 

To View Stored Readings: 

1.

 

Press and Release 

2

ND

 

2.

 

Press and Release 

ON ZERO

 

3.

 

Press and Release 

HOLD

. DATA Icon Appears on LCD 

4.

 

Press and Release 

move

 (2

ND

)

 

5.

 

Repeat step 4 to scroll readings. The newest to oldest reading is 

displayed. 

To Clear Stored Readings: Press and Hold 

ON ZERO

 

To Exit Stored Readings: Press and Release 

DATA (HOLD) 

 

3RD BUTTON FUNCTIONS 

LOCK

 

Locks Buttons/Functions/Settings. 3 button push can disable / enable lock. 3 digit 
lock Combination can be entered for additional security. 

a.

 

To Enable / Disable Lock 

1.

 

Press and Release

 2

ND

 

2.

 

Press and Release

 ON ZERO

 

3.

 

Press and Release

 TOL 

A Lock Icon Appears on LCD 

b.

 

To Program Lock Combination 

1.

 

Press and Release 

2

ND

 

2.

 

Press and Release 

ON ZERO

 

3.

 

Press and Hold 

TOL 

Until 000 Appears and SET is Flashing 

Use the 

move

 (2

ND

),

 

change

 (TOL)

 and 

apply

 (HOLD

) to program 3 

digit combination. 

Move

 scrolls flashing icon from left to right. 

Change

 toggles the value 

of the flashing icon. 

Apply

 enters 3 digit combination. Enable functions 

you want to use before locking durometer. Example: Using MAX 

HOLD, Turn the Max hold on before you lock. Lock function will lock 

out all functions that are not turned on. Functions that cannot be 

locked include OFF, ON, ZERO. 

Note: Master Combination 285 (If Combination is Lost) 

 

 

 

 

 

 