6.3.2 Installation

Field installed systems are supplied with complete installation 
and operating instructions. If your system is factory installed, the 
freezer is shipped with a coiled length of tubing to connect the 
freezer to the bottles:

3/8 in. OD copper tubing for connection to the CO

2

 supply.

5/8 in. OD copper tubing covered with Armaflex

 insulating 

tubing for connection to the LN

supply.

Straighten the coiled tubing and connect one end to the labeled 
connection on the freezer and the other end to the supply bottle or 

building supply fitting. 

6.3.3 Operation

To activate the optional backup system, complete the following 

steps:

1. Push the Power Switch to the ON position. The digital 

temperature display window shows the current cabinet 

temperature.

2. Press and hold the Push to Set button to display the setpoint 

temperature in the digital temperature display window.

3. Use a small screwdriver to rotate the Setpoint Adjust screw 

until the correct backup operating temperature shows in the 
display window. The recommended setpoint is 10 to 15

°

above normal cabinet temperature.

Note:

The temperature shown in the display window is 
accurate to 

±

1

°

C.

The backup system can run for a minimum of eight hours on 

battery power.

WARNING!

You must push the Power Switch to the 

ON position in order to activate the backup system.

6.3.4 System Status Lights

In addition to the backup control system, the built-in backup 

system includes main control panel system status lights. These 
lights are only available when the backup system is installed.

The Supply On Indicator lights when the backup system is 

operating on battery power.

The Backup System Battery Low Indicator lights when the 

backup system battery charge is low. The built-in battery 
charger recharges the battery to full charge when power is 
restored to the system.

Note:

Since rechargeable batteries degrade over time, the 
battery should be replaced after approximately three 
years.

The Backup System Empty Indicator lights when the standby 
CO

2

 or LN

2

 bottle is empty.

Figure 4.  Location of CO

2

 Connection Site, 

 

Chest Models Only

Table 1. 

Backup System Flow Rates – Cryogenic Freezers

Lbs/hr @ 75

°

Ambient Temperature 

(Flow Rate)

Center Air 
Temperature

-150

°

C

-140

°

C

-130

°

C

-120

°

C

Liquid LN

2

Chest Freezers 

5 ft

3

Empty

11.0

9.5

8.4

7.3

1/2 Full

9.3

8.0

7.0

6.1

Full

6.5

5.9

5.1

4.5

7 ft

3

Empty

11.3

10.0

9.0

8.0

1/2 Full

9.7

8.6

7.6

6.8

Full

7.8

6.8

6.0

5.2

10 ft

3

Empty

15.2

13.5

11.9

10.3

1/2 Full

12.0

10.4

9.4

8.3

Full

8.9

7.8

7.0

6.1

24.75

"

2.375

"

CO

2

 Inlet Conn.

Copper Tubing

Factory Supplie

d

0.25" O.D

.

MACHINE COMPARTMENT

REAR VIEW

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for Legaci ULT1786-9-D30

Page 1: ...underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Demos In stock Ready to ship TAR certified secure asset solutions Expert team I Trust guarantee I 100 satisfaction All trademarks brand names and brands appearing herein are the property of their respective o...

Page 2: ...egaci 1 Safety Considerations 1 General Recommendations 1 Pre Installation 1 Installation 2 Operation Ultima and Elite Freezers 4 Operation Value Freezers 9 Maintenance and Troubleshooting 10 Chart Recorders 11 Optional Equipment 12 Accessories 12 Warranty 16 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 3: ...Control Panel Features 4 6 2 Start Up 5 6 2 1 Turning the Power On 5 6 2 2 Setting the Cabinet Temperature 5 6 2 3 Setting the Alarms 5 6 2 4 Alarm Reset Status Lights Ultima II and Elite Freezers 5 6 2 5 Alarm Test Ultima II and Elite Freezers 6 6 3 Backup System Optional 6 6 3 1 CO2 and LN2 Precautions 6 6 3 2 Installation 7 6 3 3 Operation 7 6 3 4 System Status Lights 7 7 Operation Value Freeze...

Page 4: ...tenance 11 9 5 Troubleshooting Procedures 11 10 Chart Recorders 12 10 1Set Up and Operation 12 10 2Power Supply 12 10 3Changing Chart Paper 12 10 4Calibration Adjustment 12 11 Optional Equipment 13 11 1Backup Protection Devices 13 11 1 1 Liquid Nitrogen LN2 13 11 1 2 Liquid CO2 13 11 1 3 Voltage Safeguard 13 11 2Remote Alarm 13 12 Accessories 13 13 Warranty 17 13 1Domestic Warranty 13 Months Full ...

Page 5: ... 90 F ambient environment only when the freezer is used for storage WARNING This unit is not a rapid freeze device Freezing large quantities of liquid or high water content items will temporarily increase the chamber temperature and will cause the compressors to operate for a prolonged time period Attempting to utilize this freezer improperly may jeopardize safety or cause undue stress or damage t...

Page 6: ... to the National Electrical Code and local codes Do not connect the unit to overloaded power lines Your freezer is equipped with one of five NEMA style plugs refer to Figure 1 These plugs meet UL requirements Figure 1 NEMA Style Plugs Always connect the freezer to a dedicated separate circuit Each freezer is equipped with a service cord and plug designed to connect it to a power outlet which deliv...

Page 7: ...ges fully with the strike can result in substantial prying forces on the door 3 Keep applying gentle pressure to the latch handle until the latch is securely in closed position 4 Insert the key and rotate clockwise to lock 5 Replace the padlock if you are using one 5 5 Pressure Equalization Port When an upright ultra low temperature freezer door is opened room temperature air rushes into the stora...

Page 8: ...ronic panel is blown by a power surge refer to Section 8 on page 10 7 Alarm Test Pad Press this pad to start an alarm test refer to Section 5 2 5 on page 6 8 Alarm Reset Pad Press this pad to reset slowly blinking indicators Slowly blinking indicators denote that the condition has occurred but is now within the given operating parameters This pad is also depressed to temporarily silence the audibl...

Page 9: ...Alarm value 3 Press and hold or to adjust the Warm Alarm Setpoint 4 Release both pads when the digital temperature display window shows the correct setpoint value When the cabinet temperature drops below the Warm Alarm setting turn the key switch to ALARM ON The freezer is ready to operate Note If a power failure lasting over 30 seconds occurs the POWER ON light goes out the POWER FAILURE light an...

Page 10: ... Power Switch Temperature Control Setpoint Screw Temperature Control Set Button Digital Temperature Display Calibration Pot 6 3 1 CO2 and LN2 Precautions The following are precautions for using liquid CO2 and LN2 backup systems WARNING If a CO2 or LN2 cylinder falls and a valve is knocked off the cylinder becomes a deadly and completely unguided missile Transport the cylinders in a handtruck or ca...

Page 11: ...urs on battery power WARNING You must push the Power Switch to the ON position in order to activate the backup system 6 3 4 System Status Lights In addition to the backup control system the built in backup system includes main control panel system status lights These lights are only available when the backup system is installed The Supply On Indicator lights when the backup system is operating on ...

Page 12: ... 1 2 Full 4 5 5 2 5 9 Full 3 5 4 2 4 8 Full 3 4 3 9 4 4 17 ft3 Empty 6 5 7 5 8 7 17 ft3 Empty 6 2 7 1 8 1 1 2 Full 5 5 6 3 7 3 1 2 Full 5 3 6 0 6 8 Full 4 1 4 8 5 5 Full 3 9 4 4 5 1 20 ft3 Empty 7 3 8 4 9 8 20 ft3 Empty 7 0 8 0 9 1 1 2 Full 6 2 7 1 8 2 1 2 Full 5 9 6 7 7 6 Full 4 6 5 4 6 2 Full 4 4 5 0 5 7 Upright Freezers Upright Freezers 13 ft3 Empty 7 4 8 5 9 7 13 ft3 Empty 7 0 7 9 8 9 1 2 Full...

Page 13: ... the slotted screw labeled Control Setpoint and simultaneously press and hold the Setpoint Display button The temperature display changes to read the existing setpoint value 2 Turn the setpoint screw clockwise for a colder setting and counterclockwise for a warmer setting until the desired setpoint shows in the digital temperature display 3 Release the Setpoint Display button The digital temperatu...

Page 14: ...st the equipment complete the following steps 1 Remove all products and place in another cabinet 2 Turn off the unit and allow the interior to warm to room temperature 3 Dispose of the ice and wipe out any water standing in the bottom of the cabinet If there is freezer odor wash the interior with a solution of baking soda and warm water Clean the exterior with any common household cleaning wax 8 4...

Page 15: ...ttery is low the LED flashes to indicate that the battery needs to be changed 9 3 Changing Chart Paper To change the chart paper complete the following steps 1 Locate the pressure sensitive buttons at the front upper left of the recorder panel 2 Press and hold the Change Chart button 3 for one second The pen will move off the scale 3 Unscrew the center nut remove the old chart paper and install ne...

Page 16: ...ted together increasing the effective length of standby refrigeration 10 1 3 Voltage Safeguard This fully automatic device monitors the supply voltage and increases voltage if necessary so the freezer always operates within the factory recommended limits Ultima II models are factory equipped with this device CAUTION The Voltage Safeguard unit does not allow you to operate a 115 volt product on 230...

Page 17: ...ft 6098 Single split rack holds 6 2 in boxes for all chest models except 3 1 and 4 5 cu ft 6230 Single split rack holds 4 3 in boxes for all chest models except 3 1 and 4 5 cu ft 6231 Basket Rail Kit for 6 8 cu ft chest freezers 6217 Basket Rail Kit for 10 3 cu ft chest freezers 6218 Basket Rail Kit for 13 7 cu ft chest freezers 6190 Basket Rail Kit for 17 1 cu ft chest freezers 6191 Basket Rail K...

Page 18: ...ectrical spikes temporary high voltage condition Built in surge protection is standard on Revco Ultima II models free standing surge protection is optional on all other models Description Catalog No Temperature recorder 6 in circular chart seven day drive panel mounted Factory installed 6171 Temperature recorder 6 in circular chart seven day drive for 120 C 140 C and 150 C models only panel mounte...

Page 19: ...model number and voltage when ordering 5612 Deluxe Electronic Remote Alarm System User programmable to sound alarm in the event of temperature rise or power failure Can dial up to four telephone numbers to advise of alarm condition across any telephone system which accepts pulse dialing One system can monitor up to four individual freezers or up to three groups of freezers Contact Revco for detail...

Page 20: ...as set forth in this warranty and any claims for incidental or consequential damages are expressly excluded 13 2 International Warranty 16 Month Parts Warranty Revco Scientific Inc warrants this product to the owner for a period of sixteen 16 months from the date of shipment by revco If any part is found to contain a manufacturing defect in material or workmanship Revco assumes no responsibility f...

Page 21: ...ble please provide the date purchased the source of purchase Revco or specific agent rep organization and purchase order number IF YOU NEED ASSISTANCE REVCO SALES DIVISION Phone 828 658 2711 800 252 7100 FAX 828 645 3368 LABORATORY PARTS and SERVICE Phone 800 438 4851 FAX 828 658 2576 TECHNICAL SUPPORT Phone 800 438 4851 275 Aiken Road Asheville NC 28804 U S A Artisan Technology Group Quality Inst...

Page 22: ...quipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentation Artisan Scientific Corporation dba Artisan Technology Group is not an affiliate representative or authorized distributor for any manufacturer listed herein We re here to make your life easier How...

Reviews: