MAINTENANCE
It is advisable to clean the stretcher after each patient, if their is a risk of cross infection.
Do not use abrasive or solvents goods (scouring cream).
Do not use a jet of water.
Do not use alcohol products for disinfection.
RECHARGING OF BATTERY (ELECTRIC VERSION)
The batteries have to be recharged every day.
Connect the charger (supplied with the hoist) to the recharging plug located on the left side of the cover.
Recharging time : 6 hours
Recharging can be made with or without the emergency stop button engaged.
See that the charger be protected from eventual shocks.
SAFETY SYSTEMS (ELECTRIC VERSION)
The electric chair lift is equipped with two safety systems :
An emergency stop button avoiding the chair to go up and down.
An emergency lowering button: that's an independent system, hand operated, located on the frame cover.
This system allows to take the chair down by pressing continually on the button in case of bad functioning of the lift.
This function doesn't have to be used when the lift works correctly.
CHANGE OF FUSES (ELECTRIC VERSION)
CHANGE OF THE GENERAL FUSE :
Take out the defective fuse (on the right side of the cover, on the frame).
Take the new fuse (located next to the other fuse).
Replace the fuse.
It is recommended to replace the spare fuse (10A fuse).
CHANGE OF THE RECHARGING FUSE :
Do it in the same way to change the recharging fuse (1.6A).It is located on the cover.
The 1.6A recharging fuse is supplied with the lift in a pouch.
7/9