Transport, scope of delivery, installation
11
NOTICE
Atmospheric humidity
–
Electronic and mechanical components may be damaged and the
performance data alter to an unknown extent.
•
Do not exceed the admissible range for atmospheric humidity.
3.5 Removing the transport safeguard
Remove the cardboard cross, which serves to protect the knives during transport),
from the inside of the grinding jar.
3.6 Installation of the machine
Installation height: maximum 2000 m above sea level
NOTICE
Installation of the machine
–
It must be possible to disconnet the machine from the mains at any time.
•
Install the machine such that the connection for the mains cable is
easily accessible.
3.7 Electrical connection
WARNING
When connecting the power cable to the mains supply, use an external fusethat
complies with the regulations applicable to the place of installation .
•
Please check the type plate for details on the necessary voltage and
frequency for the device.
•
Make sure the levels agree with the existing mains power supply.
•
Use the supplied connection cable to connect the device to the mains power
supply.
3.8 Type plate description