iBreeze CPAP Clinical Guide
20
8 Device Alert
There are 4 types of alerts described here:
Alert 0
: Gray bottom white tips, no light, no sound, always display, disappear until the
prompt condition not met.
Alert 1
: Orange bottom white tips, blue-ray flashing, with sound, always display, disappear
until the prompt condition not met.
Alert 2
: Red bottom white tips, blue-ray flashing, with sound, exclusive alarm until the user
opens the alert to confirm or press the Therapy On/Off key.
Notification
:
White reminder message, no light, no sound, after the corresponding process,
the message automatically disappears.
Alert Summary Table: The following table summarizes the alerts.
Alert
Type
Possible Cause
Action
Input voltage is
abnormal, please
check!
Alert 2
Power Adapter with the
wrong type, resulting in
voltage is too high or too
low
Use the Power Adapter
provided by Resvent.
SD card write/read
underway, do not
remove the SD card, do
not cut off the power.
Notification
1. Insert SD card during
data synchronization;
2. Input the configuration
on the SD card.
No action.
SD Card Removed.
Notification
1. No SD card in the
device.
2. SD card has been
removed.
Reinsert functional SD
card.
SD Card Full, Please
replace the SD Card.
Notification
In Standby mode, SD card
storage space only 200M
Replace the SD card
or clean the data after
export the data in the
SD card.
SD card can’t be
written, please unlock
and insert again.
Notification
The SD card is
read-
only and can’t be
written.
Remove the SD card,
unlock and insert it
again.
SD card error, please
remove and insert
again.
Notification
SD card failure, may be:
1.SD card can’t read and
write.
2.SD card read and write
data errors.
Remove SD Card,
Reinsert or replace
with a new card.
Software update
underway, Do not cut
off the power!
Notification
Software update.
No action.
Water level is too low,
please add to the
appropriate water level.
Alert 0
During the heating
process, the water level is
below the defined
threshold.
Remove the water tank
and add to the
appropriate water
level.
System error code:
XXXX Please try to
restart, please contact
the supplier if repeat.
Alert 2
1. Pressure sensor failure,
flow sensor failure, blower
failure in therapy state; 2.
Power board short-circuit
3. Humidifier heating
circuit short-circuit
Please try to restart,
please contact the
supplier if repeat.