Restek Q-sep 3000 Operators & Service Manual Download Page 10

visit 

www.restek.com/QuEChERS

10

Safety Features 

Lid Safety Switch:

The Q-sep™ 3000 centrifuge lid is secured to the top of the cabinet by a latching knob and pawl system.

When the knob is rotated clockwise, the pawl grips the underside of the cabinet opening and prevents the lid from open-
ing. A mechanical stop positions the pawl and prevents it from rotating completely. When rotated to the stop position, the
pawl makes contact with a micro-switch mounted underneath the cabinet top. The lid safety switch prevents the cen-
trifuge from operating while the lid is open. An indicator light on the front of the machine will illuminate when the lid has
been latched properly. 

Lid Safety Interlock System:

In addition to the Lid Safety Switch, the Q-sep™ 3000 centrifuge has a true “0 rpm” lid-lock-

ing system. The lid safety interlock system keeps the lid locked at all times (even during power failure) and requires that the
rotor be at rest in order to unlock the lid. The centrifuge will not allow entry into the rotor chamber unless the centrifuge
has power and the rotor is stopped. To open the lid, make sure that the centrifuge is plugged in and, with the rotor stopped,
press the “OPEN / EMERGENCY STOP” button. 

Note: After the centrifuge has started spinning, it may be possible to rotate the lid knob enough to cause the pawl to lose contact
with the lid safety switch. If this happens, the centrifuge motor may lose power, 

but the lid will still remain locked.

If the knob

is accidentally moved and this situation should occur, rotate the knob fully clockwise to its stop position and the centrifuge will
resume operation after a few seconds. 

Circuit Breaker:

The Q-sep™ 3000 centrifuge is protected with a 4 amp circuit breaker located at the rear of the machine

and mounted to the base. Any electrical short circuit will cause the breaker to cut power to the machine. 

Imbalance Detection System:

In the event that an imbalanced load is spun in this centrifuge, the run will be aborted. A

continuous, two-count flash of the run and latch indicator lights signals that an imbalance was detected. Once the cen-
trifuge has safely come to a complete stop and unlocked, press the “OPEN” button or open the lid to reset the fault. Correct
the imbalance before proceeding. 

Emergency Rotor Chamber Entry:

In the event of power failure, it may be impossible to unlock the lid by conventional

means. In this case, entry into the rotor chamber may be made by removing the latch label and using a pen to manually
disengage the locking mechanism (see photo). Pull the mechanism towards the control panel, then unlatch and open the
lid. If the unit is damaged, contact Restek Corporation or your distributor.

Summary of Contents for Q-sep 3000

Page 1: ...s Service Manual cat 26230 26231 The Q sep 3000 centrifuge complies with all requirements of UL standard 3101 20 Can CSA C22 2 No 1010 1 and Can CSA C22 2 No 1010 2 20 NOTE The programmable run time h...

Page 2: ...maintenance records PRECAUTIONS 1 Do not attempt to unlatch the centrifuge lid unless the Unlocked indicator light is illuminated see pg 4 2 Avoid solvent contact with centrifuge lid 3 Insure that the...

Page 3: ...e centrifuge only after the rotor has completely stopped Brushless DC motor requires no routine maintenance cool running Clear lid for safe observation of samples and optical calibration of speed Elec...

Page 4: ...must be closed Open Emergency Stop Button Allows access to the rotor chamber by disengaging the locking mechanism Entry is only permitted when the rotor is stopped Pressing this button during operati...

Page 5: ...e cut to the motor but you should not be able to fully turn the latch If it is possible to turn the latch and open the lid while the unit is running contact Restek Corporation or your local distributo...

Page 6: ...l operation 1 Plug the centrifuge into an approved 115 volt cat 26230 or 230 volt cat 26231 AC 50 60 Hz outlet and press the POWER button 2 Push the OPEN EMERGENCY STOP button verify the rotor is not...

Page 7: ...perly maintained see Balanced Loads below Balanced Laods Your centrifuge must contain a balanced load in order to work properly To insure that the load is balanced keep these rules in mind when insert...

Page 8: ...spill occurs in the centrifuge or tube holder thoroughly clean the rotor and both the inside and outside of the tube holder or replace the tube holder if necessary Replacement parts can be found on p...

Page 9: ...the bottom of the rotor with the pin in the motor shaft see picture The rotor should slide onto the shaft freely and the rotor should be level Be sure that the slot in the underside of the rotor lines...

Page 10: ...started spinning it may be possible to rotate the lid knob enough to cause the pawl to lose contact with the lid safety switch If this happens the centrifuge motor may lose power but the lid will stil...

Page 11: ...authorized distributor for assistance The printed circuit board may be damaged Contact Restek Corporation or your distributor for assistance The Q sep 3000 centrifuge is equipped with severe out of ba...

Page 12: ...and workmanship for 5 000 cycles or 12 months whichever comes first Should the centrifuge require warranty or out of warranty service please contact Restek Corporation or your local distributor Lit C...

Reviews: