8.0 ProFLOW 6000 menu structure
8.1  Unit power up/power down 

 

Press the

(Power) button:

The LCD screen will display 

Followed by   

Followed by measured 

the device intro screen: 

calibrated date:  

flow data:

 

01/10/2012

 

To power off, press and hold the

(Power) button.

Other messages encountered at power up

 

If the unit is hooked up to a flow stream with a flow rate less than 0.50 mL/min, 
the unit will report an “under range” status for the flow. This message will appear 
until the flow rate exceeds 0.50 mL/min.

8.2  USB activation   

To enable the USB, press the

 

 

(Menu) button.

Use the

(arrow) keys to select the USB menu:

Press the

 

 

(menu) button again to enter the value select screen. 

Use the

(arrow) keys to toggle between USB “on” and “off.” 

To return to the main menu screen, press the

 

 

(menu) button.

To exit and return to measuring flow, press the

 

 

(menu) button again. 

 

è

 See Section 7.0: Data Collection on the PC

Summary of Contents for 22656

Page 1: ...www restek com Restek ProFLOW 6000 Electronic Flowmeter Version 5 6 Cat 22656 Instruction Manual...

Page 2: ...FLOW 6000 menu structure 9 8 1 Unit power up power down 9 8 2 USB activation 9 8 3 Adjust LCD image backlight 10 8 4 Show battery charge indicator 10 8 5 Firmware version information 11 8 6 Adjust aut...

Page 3: ...all laboratory gases CAUTION Do NOT exceed maximum operating flow rates Recalibration may be required if the unit has been subjected to extreme flow rates Always use appropriate laboratory safety pra...

Page 4: ...2 Extend the wire stand before opening battery door Figure 3 Insert the batteries as marked Type of measurement Volumetric flow Accuracy of measurements 2 of flow reading or 0 2 mL min whichever is gr...

Page 5: ...ee Section 8 4 Show Battery Charge Indicator 5 0 Operating instructions CAUTION Do not exceed maximum operating flow rates Recalibration may be required if the unit has been subjected to extreme flow...

Page 6: ...ts of mL min are equivalent to ccm 6 1 Flow range display The unit automatically adjusts the resolution of the display depending on the flow range being measured Table III shows the resolution of the...

Page 7: ...ting system for which there is an FTDIVCP driver After installing the driver connecting the device and determining which port it creates you can access the data stream through any programmatic means o...

Page 8: ...menu entry in the ProFLOW and turn the transmission on Finally click Open on the Port tab in RealTerm and you will see the serial data stream begin in the terminal window Figure 8 If you would like t...

Page 9: ...tream with a flow rate less than 0 50 mL min the unit will report an under range status for the flow This message will appear until the flow rate exceeds 0 50 mL min 8 2 USB activation To enable the U...

Page 10: ...ght Values 0 off and 5 maximum To return to the main menu screen press the Menu button To exit and return to measuring flow press the Menu button again 8 4 Show battery charge indicator Press the Menu...

Page 11: ...button again 8 6 Adjust auto shutoff duration To conserve battery life the unit automatically turns off after 6 minutes To customize the auto shutoff setting press the Menu button Use the arrow keys...

Page 12: ...0001 9 0 Troubleshooting Contact returns restek com for return instructions for servicing a damaged unit Additional charges may apply if the warranty has expired or the unit is damaged due to misuse C...

Page 13: ...S 2 compliant 10 China RoHS Logo 2 Number Font Impact China RoHS Logo 1 11 Ex nL EN60079 0 2006 Electrical apparatus for explosive gas atmospheres Part 0 General Requirements EN60079 15 2005 Electrica...

Page 14: ...w gas where the gas is at elevated temperatures the error due to humidity contributions can be extreme For the most accurate measurement of laboratory gas flow rates we recommend using the Restek ProF...

Page 15: ...n 22 8 2 USB Aktivierung 22 8 3 Die LCD Hintergrundbeleuchtung einstellen 23 8 4 Den Batterieladungsindikator anzeigen 23 8 5 Firmware Version Information 24 8 6 Die Selbstausschaltdauer einstellen 24...

Page 16: ...TEN AUSGESETZTWURDE BEI BENUTZUNG DIESES GER TS IMMER AUSREICHENDE SICHERHEITSVORKEHRUNGENTREFFEN TRAGEN SIE LABOR SCHUTZBRILLENWENN SIE DIESES GER T BENUTZEN ANMERKUNG Es gibt keine wartungsbed rftig...

Page 17: ...ist empfohlen Zertifikate CE Ex unter Sektion 10 0 Befolgungen WEEE RoHS 2 China RoHS 2 3 0 Die Batterien einbauen Dieses Ger t benutzt 2 AA Alkali Batterien Um die Batterie einzubauen muss der Draht...

Page 18: ...erieladungsindikator ausger stet Wechseln Sie die Batterie wenn angedeutet aus Sektion 8 4 Batterieladungsindikator 5 0 Bedienungsanleitung VORSICHT AUF KEINEN FALL DIE MAXIMAL ZUL SSIGE FLUSSRATE BER...

Page 19: ...ss 515 mL min bertrifft zeigt die Anzeige over range an berm ig hohe Flussraten k nnen das Ger t besch digen ANMERKUNG Die Einheit mL min ist gleichwertig mit ccm 6 1 Flussbereichsanzeige Das Ger t ju...

Page 20: ...t mit jedem Operating System f r den es einen FTDI VCP Driver gibt Nachdem der Driver installiert ist verbinden Sie das Ger t mit dem PC und stellen Sie fest welchen Port es geschaffen hat Jetzt haben...

Page 21: ...LOW und schalten Sie die bertragung ein Zuletzt klicken Sie Open am PortTab in RealTerm und Sie werden sehen wie der Serial Datenstrom imTerminalWindow beginnt Abbildung 8 Wenn Sie die Flussdaten in e...

Page 22: ...n ist zeigt das Ger t an dass der Fluss unterhalb des Messbereiches ist Diese Mitteilung bleibt auf dem Bildschirm bis die Flussrate 0 50 mL min erreicht 8 2 USB Aktivierung Um den USB einzuschalten d...

Page 23: ...leuchtungswerte 0 aus und 5 maximum Um wieder auf das Hauptmen zur ckzukommen dr cken Sie die menu Taste Um zur Flussmessung zur ckzukehren dr cken Sie die menu Taste noch einmal 8 4 Den Batterieladun...

Page 24: ...m Batterien zu sparen hat das Ger t eine Selbstausschaltung Die Selbstausschaltzeit ist auf 6 Minuten eingestellt Sie k nnen das aber auch umstellen Dazu dr cken Sie die menu Taste Benutzen Sie die Pf...

Page 25: ...en Sie returns restek com H ndler f r Anweisungen zum Service von besch digten Ger ten Es kann zus tzliche Geb hren geben falls die Garantie abgelaufen ist oder das Ger t durch Missbrauch besch digt w...

Page 26: ...Regeln 10 China RoHS Logo 2 Number Font Impact China RoHS Logo 1 11 Ex nL EN60079 0 2006 Elektrische Apparate f r explosive Gasatmosph ren Teil 0 Allgemeine Anforderungen EN60079 15 2005 Elektrische A...

Page 27: ...ch Variationen in der Luftfeuchtigkeit und der direkte Einfluss auf die gemessene Flussrate Wenn ein Bl schen Flowmeter bei erh hter Temperatur benutzt wird kann der Fehler aufgrund der Luftfeuchtigke...

Page 28: ...28 www restek com...

Page 29: ...www restek com 29...

Page 30: ...30 www restek com...

Page 31: ...w A3 Class A ENG1010 1 2001 EN60079 0 2006 EN60079 15 2005 Declarer s Name Restek Corporation Declarer s Address 110 Benner Circle Bellefonte PA 16823 Type of Equipment ProFLOW 6000 Series Flow Meter...

Page 32: ...earethepropertyoftheirrespectiveowners Restekregisteredtrademarks areregisteredintheU S andmayalsoberegisteredinothercountries Questions about this or any other Restek product Contact us or your local...