NOTE
If there is more than 1 level of LQC defined for an Assay, and LQC has
expired for this Assay, the LQC timer is reset to the time at which the
earliest LQC level passes in the set (e.g., if 2 levels of LQC are required
and Level 2 passes LQC but Level 1 fails, Level 1 must be repeated and
pass). The ‘Time to Expiry’ column in the LQC menu will display the
count down from the time when Level 2 passed.
An IQC Test runs automatically every time the RAMP® 200 is powered on. IQC can also be initiated
manually or configured to run automatically at predefined intervals using the CM
See
Edit Settings
or see
Editing Settings Using the RCU
sections.
1
From the
Main Menu
screen (
Figure 6
), select
Run IQC
. The
IQC Status
screen is displayed
(
Figure 14
).
IQC is started on all idle ports and will automatically start on other ports as they become free.
Figure 14
IQC Status Screen
2
Touch
OK
to return to the Main Menu. If IQC fails, the TM port LED illuminates red, the port
icon turns red
and access to the failed port is prevented.
NOTE
To schedule IQC Tests to run at predefined intervals, the Test times
must be configured using the CM.
Refer to Timer Tab section in
View Screen Tab
settings.