1
Hold the card so that the arrow is on the top side and pointing away from the user.
2
Insert the Lot Card – contact end first – into the slot located on the front of the CM.
3
The display provides the following status messages:
Reading Lot Card
Complete
Lot ID: M0
XXXX
Remove Lot Card
Note: The Lot ID displayed is the lot number of the RAMP® Test Kit.
4
Remove the Lot Card and store in its original pouch.
If the screen displays an error message, or if there is no response to the Lot Card
insertion, refer to the
Troubleshooting
section.
5
Keep the Lot Card for the life of the Test Kit.
NOTE
The default setting of the RAMP® 200 does not require an Operator ID
or Password. These requirements can be edited via
Edit Operator
in
the Settings menu or the RCU. If using the RAMP® 200 with default
settings,
Figure 7
will not appear.
If the following Liquid Quality Control warnings
or
appear:
Refer to
LQC icons
in
Quality Control
section for instruction on how to proceed.
NOTE
If a Test Cartridge is inserted and the Lot Card data is not stored in the
CM, the
Lot # XXXX Not Found
,
Insert Lot Card or Cancel
message will
be displayed. The Lot Card must be inserted within 40 seconds or the
Assay is aborted. Select
OK
to return to the Home screen.
1
From the
Main Menu
screen (
Figure 6
) select
Run Assay
. If prompted, enter Operator ID
and/or Password (
Figure 7
) using either the touchscreen or the barcode scanner. To accept
the displayed information, touch
Accept
.
Summary of Contents for RAMP 200
Page 7: ...A ...
Page 13: ...B ...
Page 15: ...C ...
Page 20: ...Figure 4 RAMP 200 Test Module rear panel label ...
Page 21: ...D ...
Page 25: ...E ...
Page 31: ...F ...
Page 37: ...G ...
Page 44: ...H ...
Page 54: ......
Page 55: ...I ...
Page 62: ...J ...
Page 65: ...K ...
Page 73: ...L ...
Page 85: ...M ...
Page 94: ......