Chapter 4: System Setup Procedures
Tranquility Quest Plus CPAP System Service Manual
7
1003740
Chapter 4: System Setup
Procedures
NOTE:
This section provides directions for accessing the
Therapy Menu. Prescribed therapy settings can
only be set using the Therapy Menu. (To prevent
patients from tampering with the settings, the
directions to access the Therapy Menu should not
be revealed to the patient.)
CAUTION:
If the Tranquility Quest Plus CPAP System has
been exposed to either very hot or very cold
temperatures, allow it to adjust to room
temperature (approximately one hour) before
beginning setup.
4.1 Pressure and Ramp Settings
The pressure and ramp settings for the unit are set using the
Control Panel on the bottom of the unit. When not in use the
Control Panel display will show the current compliance meter
reading for the unit. This reading will reflect the total number of
hours that the patient has been breathing on the unit.
With the unit turned on and looking at the bottom, press the up
and down arrow button simultaneously for three seconds and the
pressure and ramp setting will be displayed. The Minimum
Pressure setting will now be flashing. Press the up or down arrow
button to change the setting. Press the mode button to move to the
ramp setting. Press the up or down arrow button to change the
setting. The setting is implemented immediately when the up or
down button is pressed. Press the mode button for three seconds
to lock the controls.
The display will revert to the compliance meter display 60 seconds
after the last time a button has been pushed. The settings can only
be changed if they are flashing, otherwise they are locked.