background image

14

Disposal

Dispose of this device in accordance with local regulations.

How to Contact Respironics

To have your device serviced, contact your home care provider. If you need to contact Respironics directly, call the 
Customer Service department at 1-800-345-6443 or 1-724-387-4000.
You can also use the following address: 
Respironics, Inc.
1001 Murry Ridge Lane
Murrysville, PA 15668

Summary of Contents for System One Heated Humidifier

Page 1: ...System One Heated Humidifier USER MANUAL ...

Page 2: ... 2009 Respironics Inc and its affiliates All rights reserved ...

Page 3: ...rapy Device 4 Disconnecting the Therapy Device 5 Daily Use 5 Home Cleaning Water Tank 9 Home Cleaning Humidifier Base 9 Hospital and Institution Cleaning 10 Service 10 Traveling with the System 10 Troubleshooting 11 Specifications 13 Disposal 14 How to Contact Respironics 14 EMC Information 15 Limited Warranty Back Page ...

Page 4: ...efore cleaning the humidifier always remove from the therapy device Empty and clean the water tank daily to prevent mold and bacteria growth Wipe the seal completely Note Please see the Limited Warranty section of this manual for information on warranty coverage Cautions A caution indicates the possibility of damage to the device Do not place the humidifier directly onto carpet fabric or other fla...

Page 5: ...Port Connects to the outlet port on the therapy device Air Outlet Port conical 22 mm Connect the patient circuit here Heater Plate Warms the water in the water tank Humidifier Door Open the door to access the water tank Humidifier Door Lever Lift up on the lever to open the humidifier door Water Tank The removable water tank holds the water for humidification Maximum Fill Lines The fill lines indi...

Page 6: ...y from the unit Therapy Device Side Cover Tab 3 Next line up the components side by side Make sure that the guides on the humidifier fit into the slots on the therapy device and the air outlet port on the therapy device fits into the air inlet port on the humidifier Therapy Device Humidifier Air Inlet Port Air Outlet Port Slots Guides 4 Simply press the two units together until they snap into plac...

Page 7: ...precautions to protect furniture from water damage CAUTION Do not turn the humidifier on without the water tank installed The Humidifier number setting must remain on 0 if there is no water in the water tank CAUTION Do not attempt to fill the tank while it is still inside the humidifier WARNING Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removing th...

Page 8: ...ml no higher than the maximum fill line located on the front and both sides of the tank Tank Base Tank Lid MaximumWater Fill Lines CAUTION Use only room temperature distilled water in the tank Do not put any chemicals or additives into the water Possible airway irritation or damage to the water tank may result CAUTION Do not fill the water tank above the maximum fill line Damage to the humidifier ...

Page 9: ...Do not move the humidifier while the water tank has water in it 7 Plug the power supply cord into the back of the therapy device 8 Plug the AC power cord into the power supply provided with your therapy device and then plug the AC power cord into an electrical outlet that is not controlled by a wall switch 9 Attach the flexible tubing from the patient circuit to the humidifier s air outlet port IM...

Page 10: ...eases CAUTION Do not turn the humidifier on without the water tank installed The Humidifier number setting must remain on 0 if there is no water in the water tank IMPORTANT The Humidifier LED will remain on but will dim when the Humidifier is attached The number may turn off depending on the LED Backlight settings IMPORTANT The ideal humidity setting depends on room temperature and humidity Initia...

Page 11: ...r use the water tank if the tank lid does not fit comfortably on the tank base 6 Before using the tank fill it with distilled water per the directions in the Daily Use section no higher than the maximum fill line located on the front and sides of the tank 7 Reassemble the tank by placing the hinges on the tank lid over the 2 tabs on the back of the tank base Close the lid until the tab on the lid ...

Page 12: ...e Discard and replace any damaged parts Service The humidifier does not require routine servicing If any part of the humidifier is worn or damaged contact Respironics or your home care provider See the Troubleshooting section later in the manual for additional information Traveling with the System Packing the System 1 Remove the water tank and empty all water 2 Put the empty water tank back into t...

Page 13: ...eated in the humidifier Remove the water tank from the humidifier base and then slide the tank back in the humidifier making sure it is pushed back as far as it can go If high leak persists issue may be caused by a misaligned or damaged seal Seals are not user serviceable components Please contact your home care provider or Respironics The humidifier LED icon on the therapy device is flashing Ther...

Page 14: ...o apply pressure until the door snaps past the locked open position Verify that the door is working properly If further assistance is needed contact your home care provider The humidifier has fallen off your table or night stand The humidifier may not have been properly seated on the night stand or the placement of the tubing may have caused the device to fall Always make sure your humidifier is p...

Page 15: ...trical Equipment EN ISO 8185 2007 General Requirements for Humidification Systems Electrical When the heated humidifier is used with a Respironics therapy device AC Power Consumption 100 240 VAC 50 60 Hz 2 1 A DC Power Consumption 12 VDC 5 0 A Type of Protection Against Electric Shock Class II Equipment Degree of Protection Against Electric Shock Type BF Applied Part Degree of Protection against I...

Page 16: ... Respironics To have your device serviced contact your home care provider If you need to contact Respironics directly call the Customer Service department at 1 800 345 6443 or 1 724 387 4000 You can also use the following address Respironics Inc 1001 Murry Ridge Lane Murrysville PA 15668 ...

Page 17: ...arge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical Fast Transient Burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for supply mains 1 kV for input output lines Mains power quality should be that of a typi...

Page 18: ...ss the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above the device should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or reloca...

Page 19: ...s to which no problem is found after investigation by Respironics Inc Service Respironics Inc disclaims all liability for economic loss loss of profits overhead or consequential damages which may be claimed to arise from any sale or use of this product Some states do not allow the exclusion or limitation of incidental or consequential damages so the above limitation or exclusion may not apply to y...

Page 20: ...1058288 1057395 JR 5 29 2009 EN DOM Manufactured for Respironics Inc 1001 Murry Ridge Lane Murrysville PA 15668 USA ...

Reviews: