Chapter 5
Repair
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
5-21
7. Return the hardware screen to its initialized state.
Pressure Relief Valve
1. Set the analyzer to read cmH
2
O.
2. Connect the analyzer’s positive low pressure port to the patient circuit
Figure 5-13: Pressure Relief Valve Troubleshooting Setup
3. Touch
Safety
to energize (white background) the safety solenoid.
4. Touch
Exhalation
and set it to 2000 steps.
5. Touch
Air
and set the flow to 1 LPM.
6. Check that the analyzer reads 130-140 cmH
2
O. If not, adjust the
relief pressure (Figure 5-14):