Component Removal/Installation
12. Connect all of the interconnect cables to the 10.4-in. GUI and
reinstall the potentiometer knobs with the set screws toward the face
of the GUI (on 9.4-in. GUIs the knobs were installed with set screws
toward the inside of the enclosure). Ensure that the knobs don’t bind
or rub when turned completely in either direction.
13. Align the foil shield holes with the power supply mounting holes.
Reinstall the power supply. Reconnect J2 and J3 harness connections.
14. Reattach the power supply shroud. Remove the backlight inverter PCB
from the power supply shroud and discard.
15. Reattach the sensor PCB (see “Sensor PCB” on page 9-5) and
reconnect all remaining interconnect cables.
Installing the CPU PCB
Remove the existing CPU PCB and replace with the CPU PCB included in the
upgrade kit. Secnum describes how to install the CPU PCB.
Downloading Ventilator Software
See Chapter 7 for complete software download instructions.
1. Connect the service PC to the ventilator serial port using a null modem
cable.
2. Insert the field upgrade CD-ROM and follow the onscreen instructions.
Use this command to prompt for the ventilator serial number (
x:
is the
CD-ROM drive, and a space precedes
-vs
):
(x:)\setup -vs
Calibrating the Screen
1. Cycle ventilator power while holding down the
Exp. Hold
and left
Options
button.
2. The ventilator prompts you to calibrate the screen. Upon
acknowledgement, the software automatically detects and adjusts for
the 10.4-in. screen size.
3. After the ventilator confirms the screen calibration is complete, cycle
power to the ventilator and enter diagnostic mode.
4. Enter the software screen and confirm that the serial number is
correct.