Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-25
Figure 9-27: Front of Ventilator with 10.4-in. GUI Removed
Intensity and Volume
Potentiometers
(9.5-in. GUI)
Follow these steps to remove the intensity and volume potentiometers (Figure
9-28). Reverse removal steps to install as shown.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Remove the sensor PCB (see “Sensor PCB” on page 9-5).
3. Remove the power supply (see “Power Supply Fan/Shroud” on page 9-
7).
4. Remove the MMI PCB (see “MMI PCB (9.5-in. GUI)” on page 9-15).
5. Remove the GUI assembly (see “GUI Assembly (9.5-in. GUI)” on
6. Remove the nut and washer holding each potentiometer, then
disconnect the harnesses and remove the potentiometers.