User Manual
37
C
HAPTER
11: S
PECIFICATIONS
E
NVIRONMENTAL
Operating
Storage
Temperature
5ºC to 35ºC
-20ºC to 60ºC
Relative Humidity
15 to 95%
(non-condensing)
15 to 95%
(non-condensing)
Atmospheric Pressure
(5600 feet to sea level)
83 to 102kPa
P
HYSICAL
Dimensions:
24.8 cm L x 16.8 cm W x 11.2 cm H
(9.75 in. L x 6.625 in. W x 4.4 in. H)
Weight:
1.8 Kilograms
E
LECTRICAL
AC Voltage Source:
100 to 240 VAC, 50/60 Hz
DC Voltage Source:
12 VDC (when operated with the external DC
power adaptor accessory)
AC Current:
1.25 A maximum
DC Current:
3.0 A maximum
Protection against electric shock:
Class II
Degree of protection against electric
shock:
Type BF Applied Part
Degree of protection against harmful ingress of water:
BiPAP autoSV Advanced device
:
Ordinary Equipment, IPX0
AC Power Supply
(Reorder number 1012832):
Drip Proof, IPX1
DC Power Adapter
(Reorder number 1012975):
Drip Proof, IPX1
Modes of Operation:
Continuous
Electromagnetic
Compatibility:
Th
e BiPAP autoSV Advanced device meets the
requirements of EN 60601-1-2, second edition
(2001).
Fuses:
Th
ere are no user-replaceable fuses.