Chapter 6
6-22
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
REF 1029568 Rev B
Ventilator Communications
6.4 Reprogramming
the BiPAP Focus
Serial Number
The serial number of each BiPAP Focus Ventilator is stored in non-volatile
memory on the user interface PCB, and must be reprogrammed if you replace
the PCB.
1. Write down the serial number from the bottom of the unit, include
hyphens if the serial number includes them. Do this
before
reprogramming the serial number.
2. All necessary software should already be installed. If not, follow the
instructions in this chapter from the beginning.
3. Enter Download mode and ensure that BiPAP Focus communication
cables are still connected to the service PC.
4. Double-click on
TUI_download_wizard.exe
on the software CD
Figure 6-40: BiPAP Focus Software CD Files