Chapter 3
3-4
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
REF 1029568 Rev B
Theory of Operation
3.2 Electronic System
The BiPAP Focus electronic system controls, monitors, and displays patient
and ventilator data. Two microcontrollers (on the user interface and controller
PCBs) interpret user and sensor inputs while controlling inspiratory and
expiratory pressures according to the selected ventilator settings. Figure 3-2
shows an interconnect block diagram of the BiPAP Focus ventilator.
Figure 3-2: BiPAP Focus Block Diagram
Communication between the microcontrollers allows each to check the
function of the other, ensures proper transition between AC and DC power
sources, and allows internal battery maintenance.
Exter
nal
AC/
DC
Motorola 68332 microcontroller,
A/D, primary alarm, temperature
monitor, primary alarm driver,
voltage monitoring,