Chapter 8
8-6
BiPAP Vision Service Manual © Respironics, Inc.
1062788
Options and Upgrades
3. Remove the 2 mounting screws that attach the flow body to the circuit
board, using care not to damage the thermistor wire connection
(Figure 8-6).
4. Align the nylon washers over the screw holes. The washers provide the
necessary clearance for installing the baffle.
Figure 8-6: Installing Nylon Washers under AFM Mounting Screws
5. Replace the flow body and fasten the screws (Figure 8-7).
Figure 8-7: Reassembled AFM
Nylon
washers (x2)
AFM mounting
screws (x2)
Flow body
Circuit board
Nylon washers (x2)
Flow body