B
I
PAP S
YNCHRONY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
S
ETUP
- P
AGE
13
1032411, R
EV
. 00
S
OFTWARE
•
Respironics Encore Pro Data Management software for reading compliance data.
DC P
OWER
A
CCESSORIES
The BiPAP Synchrony can be used in a stationary recreational vehicle, boat, or motor home when
powered by the DC/DC power adapter.
The Respironics DC/DC power adapter and cord enables the BiPAP Synchrony to be operated from a
12 VDC free-standing battery, or a portable battery pack from Respironics (RI p/n 1028869).
A
DDING
A
H
UMIDIFIER
When using a humidifier, always disconnect the humidifier tubing from the BiPAP Synchrony system
when it is turned off. DO NOT use a room humidifier within 6 ft. of the BiPAP Synchrony. Moisture can
build up in the system and cause damage. Follow the instructions included with the humidifier.
.
CAUTIONS
•
Only use the Respironics DC Power Accessories. Use of any other system may
cause damage to the BiPAP Synchrony.
•
DC Power is not intended to be used as battery back-up. DO NOT connect the
DC Power while the BiPAP Synchrony is operating on AC Power. System dam-
age may occur.