Lab
RAM II Jacketed Vessel Users’ Manual
SAFETY INFORMATION
Resodyn Acoustic Mixers, Inc.
6
Rev. 110119
WARNINGS:
Failure to properly seal the heat transfer fluid circuit
prior to starting the pump or failure to depressurize the heat
transfer fluid circuit prior to opening may result in leakage of heat
transfer fluid from the circuit. Use appropriate measures to protect
operator and equipment.
Turn heat transfer fluid pump off before removing or installing
tubing or opening the Jacketed Vessel.
CAUTIONS:
Use recommended fittings only. Use of improper
fittings and/or improper connection to fittings may result in
leakage at the fittings.
A (user supplied) heat transfer fluid circuit pressure relief/bypass
valve must be installed in a potentially high-pressure system to
prevent the pump from exceeding the maximum rated pressure;
i.e., 50 psig (3.5 bar) for the Jacketed Vessel system.
A (user supplied) product-side pressure relief valve must be
installed (i.e. in the vacuum line) if the LabRAM II Jacketed
Vessel is used in applications where the process pressure inside
the mixer could exceed 50 psig.