39
Technical Specifications
Technical Specifications
Dimensions (H x W x D)
5.6" x 9.8" x 11.5" (142 mm x 248 mm x 293 mm)
Weight
Flow generator with cover: 8.14 lb (3.7 kg)
Flow generator with Humidaire 2i (empty): 9.24 lb (4.2 kg)
Environmental conditions
Operating temperature: 41
o
F to 104
o
F (+5
o
C to +40
o
C)
Storage and transport temperature: -4
o
F to +140
o
F (-20
o
C to +60
o
C)
Humidity—operating, storage and transportation: 10%–95% non-condensing
Housing construction
Injection moulded plastic
Air tubing
1 x 6’6” (2 m)
Air filter
Synthetic fiber
Patient connection port
22 mm taper, compatible with EN 1281-1:1997 Anaesthetic & Respiratory
Equipment —Conical Connectors
Power supply
AC Input range: 110–120 V and 220–240 V; 50–60 Hz; 60 VA
DC Input range (if using DC converter): 12–13.8 V DC
Performance
Operating pressure range: 4–20 cm H
2
O
Sound pressure level
<30 dBA (tested in accordance with the requirements of ISO 17510-1)
Battery type
Lithium
Alarm specifications
•
High maximum mask pressure alarm
Stops air delivery at 25 cm H
2
O >700 ms.
•
High mean mask pressure alarm
Stops air delivery when mean mask pressure exceeds 15 cm H
2
O (averaged
over approximately 1 minute).
•
Low mask pressure alarm
Mask pressure is less than 3 cm H
2
O for longer than 5 seconds while the
patient is using the
VPAP Adapt
SV.
Summary of Contents for VPAP Adapt SV
Page 1: ...VPAP Adapt SV User Guide English Front Cover User qxp 05 04 2006 4 31 PM Page 1...
Page 3: ...VPAP Adapt SV User Guide English Reorder number 26869 2 06 04...
Page 10: ...4...
Page 12: ...6...
Page 32: ...26...
Page 36: ...30...
Page 42: ...36...
Page 44: ...38...
Page 54: ...48...