NOT013281-8
Calibrating the oxygen sensor to 21% FiO
2
This function allows you to manually calibrate the oxygen sensor to 21% by passing ambient air
through the sensor.
2 Low-pressure oxygen
Before using low-pressure oxygen, check that:
•
The oxygen sensor is installed in the ventilator (see “Installing the oxygen sensor” on
page 38)
•
The manual test was successful (which means that the oxygen sensor is working).
See “3 Manual test” on page 48
•
The “Low pressure oxygen input” parameter is set to YES (see “Low-pressure oxygen
input” on page 63).
5.11 Using the nebuliser
Before using the nebuliser, check that:
•
The oxygen sensor is installed in the ventilator (see “Installing the oxygen sensor” on
page 38)
•
The manual test was successful (which means that the oxygen sensor is working).
See “3 Manual test” on page 48
•
The high-pressure oxygen equipment is connected to the ventilator (see “High-pressure
oxygen” on page 39)
•
The “Low pressure oxygen input” parameter is set to NO (see “Low-pressure oxygen
input” on page 63).
ACV
ACV
info
HP
HP
0
10
20
30
40
50
60
70
80
90
23
c
m
H2O
M
od
es
Pa
ra
m
<
PE
<
PE
P
16
0
28
0
>
>
v
O
2
100%
2
min
Freeze->Touch curves
P
peak
peak
20,3
c
m
H
2
O
FiO
FiO
2
21
%
O
2
2
Consumed:
Consumed:
- - l/
m
in
- - l/
m
in
V
Ti
Ti
501
m
l
V
Te
Te
503
m
l
1/4
1. Press
to supply 100% FiO
2
for two minutes. The
corresponding icon will flash.
2. Press
again to interrupt the supply before the end of
the two minutes.
Note:
You cannot interrupt the oxygen supply during the first
four ventilation cycles, because this is the time required to
calibrate the sensor.
O
2
100%
2 min
O
2
100%
2 min
0
10
20
30
40
50
60
70
80
90
<
PE
00
H
2
O
cm
New configuration
Stop
m
achine
Back
Locked touch screen
MENU
MENU
Configuration
Options ...
Select a
m
enu
ACV
ACV
info
Nebul.
Nebul.
Config.
Config.
Setting
Setting
Calibr.
Calibr.
O
2
HP
0
10
20
30
40
50
60
70
80
90
<
PE
0
H
2
O
cm
Back
Ventil.
Ventil.
stop
stop
┘
HP
1. After selecting one of the configuration types (New, Most
Recent or Preset) on the main screen, press the
button on
the right-hand side of the device.
2. Press .
The following screen is displayed.
3. Press “Calibr. O
2
” to start the calibration.
MENU
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...