Introduction
NOT013281-8
3 External battery pack
Like the mains power pack, the external battery pack is housed in the compartment located on
the underside of the ventilator. When available, this power supply is used in preference to the
internal battery.
During ventilation, the battery will be recharged if the voltage is higher than 20 V. However, when
ventilation is not being delivered, the battery will be recharged whatever the voltage.
It is advisable to remove the external battery pack if the ventilator is stored for a prolonged period.
The external battery will be recharged when the ventilator is connected to:
•
The mains power pack via the external power cord
•
An external DC power supply.
When the charge of the external battery falls below 5%, the Elisée 350 switches to internal
battery power.
4 Internal battery
During ventilation, the internal battery will be recharged by an external power supply (external DC
or mains) if the voltage is higher than 20 V. However, when ventilation is not being delivered, the
battery will be recharged whatever the voltage.
2.5 Air inlet
Ambient air passes through a removable dust filter before entering the air circuit.
Top view
Underside view
Internal battery (Li-ion)
Air inlet
(
: do not cover)
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...