4

Note: 

If required, the External Battery DC output cable can be 

connected into the DC input port of the Astral device while the 

External Battery is charging. The Astral device will then continue 

to operate using mains power.
3. A fully charged External Battery is indicated by  . To 

discontinue charging, disconnect the External Battery from 

the AC power cable.

Powering on the External Battery

To power on the External Battery:

 

• Connect it to an Astral device that is powered on or;

 

• Press 

 .

Note: 

The External Battery automatically powers off after 5 

minutes of inactivity.

Using and recharging the External Battery

When connected to Astral and mains power, the External 

Battery recharges (refer to the Charging sequence section in this 

guide).

Using and recharging with the Astral device

Connecting a fully charged External Battery to the Astral device 

can provide 8 hours of power during typical use. 

Refer to illustration B.

1. Connect the External Battery DC output cable into the DC 

input port of the Astral device.

2. Connect the AC power supply cable to the DC input port of 

the External Battery.

3. Plug the other end of the AC power supply cable into the 

mains power outlet. The AC mains indicator on the Astral 

interface will illuminate.

Using two External batteries with the Astral device

A second fully charged External Battery can be connected to the 

Astral device to provide a further 8 hours of power during typical 

use. A maximum of two External batteries can be connected to 

the Astral device.

 CAUTION

Do not attempt to connect more than two external batteries. 
Battery specific messages and alarms on the Astral device 
will not operate for any additional units.
Refer to illustration C.

1. Connect the first External Battery DC output cable into the 

Astral DC input port.

2. Connect the second External Battery DC output cable into the 

DC input port of the first External Battery.

3. Connect the AC power supply cable to the DC input port of 

the second External Battery.

4. Plug the other end of the AC power supply cable into the 

mains power outlet. 

Summary of Contents for Astral External Battery

Page 1: ...User guide English...

Page 2: ...B A 1 2 3 4 5 6...

Page 3: ...eneral warnings and cautions Further specific warnings cautions and notes appear next to the relevant instruction in the user guide A warning alerts you to possible injury WARNING The battery is to be...

Page 4: ...n the battery Ensure the External Battery is kept dry Do not expose to water rain or high moisture levels Indicators Refer to illustration A The capacity and operation of the External Battery are indi...

Page 5: ...rge Maintenance displays the full charge capacity and the charge cycle count for any batteries currently detected by the system Regularly check the charge level of the external battery It is recommend...

Page 6: ...ct the External Battery DC output cable into the DC input port of the Astral device 2 Connect the AC power supply cable to the DC input port of the External Battery 3 Plug the other end of the AC powe...

Page 7: ...ttery alarm before swapping over to the other external battery The Astral device will continue to ventilate using the internal battery while the external batteries are swapped WARNING The internal bat...

Page 8: ...uld be carried out regularly Wipe the exterior of the battery with a clean slightly damp cloth Refer to the User Guides for the patient interface humidifier and other accessories in use for detailed i...

Page 9: ...ated Power connections are disrupted Check all cables and connect them as described in Setting up External Battery is discharged Connect the device to the mains power supply and recharge the External...

Page 10: ...400 charge cycles at 23 C to 80 capacity Operating conditions Charging Discharging 10 C to 35 C 5 93 maximum humidity 0 C to 40 C 5 93 maximum humidity Transport storage conditions 20 C to 50 C 5 93 r...

Page 11: ...transport External Battery charge level External Battery charging indicator Direct current O I DC on off DC Input port DC Output Do not use if package is damaged Canadian Standards Association Manufac...

Page 12: ...ices 2 years This warranty is only available to the initial consumer It is not transferable If the product fails under conditions of normal use ResMed will repair or replace at its option the defectiv...

Page 13: ...limitation may not apply to you This warranty gives you specific legal rights and you may also have other rights which vary from region to region For further information on your warranty rights conta...

Page 14: ...1 2 3 C D...

Page 15: ...de Astral is a trademark and or registered trademark of the ResMed family of companies For patent and other intellectual property information see ResMed com ip Actichlor is a trademark of Ecolab US In...

Reviews: