8
Therapy pressure
3 to 40 cm H
2
0
Dead space
information
Physical dead space (mask and elbow) using the large cushion is
366 mL.
Resistance with
Anti Asphyxia Valve
(AAV) closed to
atmosphere
Drop in pressure measured
(nominal)
at 50 L/min: 0.5 cm H
2
O
at 100 L/min: 1.1 cm H
2
O
Inspiratory and
expiratory
resistance with Anti
Asphyxia Valve
(AAV) open to
atmosphere
Inspiration at 50 L/min
1.0 cm H
2
O
Expiration at 50 L/min
1.5 cm H
2
O
Anti Asphyxia Valve
(AAV) open-to-
atmosphere
pressure
≤2 cm H
2
O
Anti Asphyxia Valve
(AAV) closed-to-
atmosphere
pressure
≤2 cm H
2
O
Summary of Contents for AcuCare F1-0
Page 1: ...Hospital masks User guide English Svenska Dansk Norsk Suomi Polski...
Page 2: ...AcuCare F1 0 AcuCare F1 1 AcuCare F1 4 1 2 4 5 3 ii...
Page 70: ...2 PPV AcuCare F1 4 AcuCare F1 4 PAP 30...
Page 71: ...3 AcuCare F1 1 AcuCare F1 4 AcuCare F1 4 22 1 2 3 4 5 1 2...
Page 72: ...4...
Page 76: ...8 ResMed com symbols...
Page 78: ...2 30 PPV AcuCare F1 4 AcuCare F1 4 PAP 30...
Page 79: ...3 AcuCare F1 1 AcuCare F1 4 AcuCare F1 4 22 1 2 3 4 5...
Page 80: ...4 1 2...
Page 81: ...5 5 40 15 95 20 60 95 148 x221 x102 158 x230 x105 165 x234 x107 www resmed com ResMed...
Page 84: ...8...