XCell™ C410:V3 Controller
User Guide
20 repligen.com UG-C410v3-06
Figure 8. Main screen
The Main screen provides an overview of the process. It displays a pneumatic process schematic, an
animated diaphragm movement /cycle, valve transition between pressure and exhaust cycles and
flow direction. From the Main screen, an operator can:
1.
Monitor and control XCell ATF® Device processes. In the Main screen and all subsequent
screens, all data fields with a white background are for display only. The operator, based
upon security levels, can change data fields with a beige background.
2.
Start/Stop diaphragm pump.
a.
When starting the diaphragm pump, a dialog box will appear to enable the user to
start with current settings, start with default settings or cancel and return to the
main screen.
b.
When stopping the diaphragm pump, a dialog box will appear that enables the user
to confirm the stop command, or to cancel and return to the main screen.
3.
Access other screens based upon password security levels.
4.
Observe P2 trending. A P2 Trend button hides/reveals this screen.
5.
Observe an animated diaphragm pump showing inflation and deflation cycles of the
diaphragm pump.
6.
Monitor Overtime condition- displayed in FLOW status sub screens, by change of actual
cycle time field to red.
7.
Monitor Overflow condition- displayed in FLOW status sub screens, by change of actual cycle
time field to orange.