REPLIGEN TangenX SIUS Cassette User Manual Download Page 2

IF.PUG.006 R10 

 

Figure 1 

Torque Sequence for TangenX™ Sius™ 

Cassettes 

PRODUCT CONTENTS  

Package includes the following: 

1.

 

TangenX™ SIUS™ membrane filtration cassette: product consists of either a 0.5 m² 

(5 ft²) cassette, 1.5 m² (15 ft²), or a 2.5 m² (25 ft²) single-use cassette 

2.

 

EPDM gaskets 

IMPORTANT INFORMATION BEFORE YOU BEGIN 

CASSETTES 

 

Product is packaged wet and must remain hydrated for optimal performance. Keep 

bag sealed until cassette installation (step 3 below). 

 

TangenX™ SIUS™ Cassettes are compatible with the TangenX™ PRO cassette 

holder. A list of other cassette holders is shown in the cassette holder compatibility 

guide. 

 

Cassettes may be stacked to increase filtration surface area; however, use only one 

type of membrane molecular weight cutoff at one time.  

Do not install a mixture of 

cassettes with different pore sizes in the hardware.

 

 

Cassettes must be equilibrated with an appropriate buffer (i.e., phosphate buffered 

saline) to ensure the neutralization of the 0.2M sodium hydroxide storage agent in 

the membrane filter.  It is important to use pre-filtered buffer to avoid fouling the 

membrane or introducing contaminants into the system that could affect membrane 

performance and product recovery. 

GASKETS 

 

Gaskets are intended to be single use; Repligen recommends that you replace 

gaskets with each cassette changeover.  Repligen supplies two gaskets per 

cassette.  Installation of the first cassette requires two gaskets; stacking additional 

cassettes requires only one gasket.  Extra gaskets should be saved to replace worn 

or damaged gaskets.

 

PUMP 

 

When using TangenX™ cassettes, select a pump with adequate capacity. Crossflow 

rate ranges (see Table 3) are feed channel type and process fluid dependent. 

TANGENX™ SIUS™ CASSETTE INSTALLATION 

1.

 

Lift the end plate off the manifold of the TangenX™ PRO cassette holder. 

2.

 

Rinse the EPDM gaskets with deionized water or WFI.  Place a rinsed gasket flat 

against the bottom manifold; ensure that the holes in the gasket line up with the 

holes in the manifold. 

3.

 

Using scissors carefully open the cassette bag to remove cassette. 

WARNING:  Each cassette is stored in 0.2M sodium hydroxide 

solution as a preservative.  Follow standard safety procedures for 

handling 0.2M sodium hydroxide, including the use of gloves, safety 

goggles, and lab coat. 

4.

 

Place the cassette into the holder flat against the gasket.  Place another gasket on 

top of the cassette.  Ensure that the holes in the manifold, gaskets, and cassette are 

completely aligned.  If you are using multiple cassettes, continue the same 

gasket/cassette/gasket pattern, ending with a gasket between the last cassette 

and the end plate. 

5.

 

Place the end plate on top of the last gasket of the cassette or cassette stack. 

6.

 

Install the tie-rod spacers (if used) and washers on each bolt leaving a minimum of 

18 mm (0.75 inch) of thread exposed on the rod.  By hand, screw the nut on each 

bolt and hand tighten evenly by alternating from one nut to the other.   

7.

 

Bolts must be further tightened to within the recommended torque values as shown in 

Table 1 using a calibrated manual torque wrench. 

 

 

 

 

 

 

 

Table 1

 Recommended Torque Values 

Holder Type 

Torque Range (in-lbs)

 

Torque Range (nm) 

TangenX™ PRO 4

-

BOLT

 

300 – 450 

35 – 50 

TangenX™ PRO 2

-

BOLT

 

600 – 900 

70 – 100 

 

8.

 

Figure 1 illustrates the bolt patterns for the two holder types: 

TangenX™ PRO 

4-bolt 

horizontal cassette holder is designated by letters H1 through H4; TangenX

™ PRO 

2-bolt vertical holder is designated by letters V1 and V2. 

9.

 

4-bolt  torque sequence:  

Using the calibrated torque 

wrench with a 1-¼" deep 

style socket, place the socket 

on bolt H1 and tighten the nut 

¼ turn.  Then move the wrench 

diagonally to bolt H2 and 

tighten the nut ¼ turn.  Next 

move the wrench to bolt H3 

and tighten the nut ¼ turn.  

Then move the wrench 

diagonally to bolt H4 and 

tighten the nut ¼ turn.  

Alternate back and forth using this crisscross pattern until the torque wrench "clicks" 

at each nut.  Repeat this sequence until the wrench "clicks" without turning the nut.  

The "click" of the torque wrench indicates that the nut has reached the set point 

torque value. 

 

10.

 

2-bolt torque sequence:  

Using the calibrated torque wrench with a 1-¼" deep 

style socket, place the socket on bolt V1 and tighten the nut ¼ turn.  Next move the 

wrench across to bolt V2 and tighten the nut ¼ turn.  Alternate back and forth until 

the torque wrench "clicks" at each nut.  Repeat this sequence until the wrench "clicks" 

without turning the nut.  The "click" of the torque wrench indicates that the nut has 

reached the set point torque value. 

CAUTION:  Nuts must be tightened uniformly to avoid damaging the 

cassette.  Leakage may result from non-parallel plate alignment or 

over-compression of the cassettes at one end.   

Wait 5-10 minutes and allow the gaskets to relax before re-torqueing.  Check each 

nut, per the Figure 1 sequence, using the torque wrench at its set point torque value 

(see step 7 above).    

11.

 

Re-torque as needed to create a liquid-tight seal, but do not exceed the maximum 

torque limit for the TangenX™ PRO holder type used (see Table 1). 

NOTE:  Torque may change during processing as the cassettes may 

compress, or as the cassettes expand or contract with temperature changes.  

Periodically check the torque of the bolts and adjust torque as needed.  

 

EQUILIBRATION OF TANGENX™ SIUS™ CASSETTES 

Cassettes must be equilibrated with an appropriate buffer (i.e., phosphate buffered 

saline) to ensure the neutralization of the 0.2M sodium hydroxide storage agent in the 

membrane filter.  Verify the pH of the effluent from the cassette is neutralized to 

minimize any possible interaction with your particular application.  For most applications, 

further sanitization is not required. 

 

 

 

 

 

 

 

INTEGRITY TEST  

 

The integrity test is a non-destructive method to verify the integrity of a TFF cassette.  

Each cassette manufactured by Repligen undergoes strict release testing, including an air 

integrity test.  This release test verifies the integrity of the cassette prior to shipment; 

however it can't guarantee the integrity of the cassette's installation in the holder at the 

time of use. In the rare case there is an integrity issue, it can be a result of shipping 

damage or improper installation. Therefore, a pre-use integrity test should be conducted 

on site and can easily be performed following the cassette installation and flush. Integrity 

test specifications are shown in Table 5 of this guide.  A detailed procedure (AN1002) 

for the measurement of air integrity can be obtained by contacting Repligen or your 

local sales representative. 

CLEANING OF THE CASSETTE SYSTEM 

TangenX™ SIUS™ cassettes are intended for single use only, post-use cleaning and re-

use is not recommended.  To clean the TFF system following use, recirculate 0.5M sodium 

hydroxide through the system with all valves open.  Cassettes are left in place during the 

system cleaning procedure to provide a flow path for the cleaning solution.  Alternatively 

the cassettes may be removed and a spacer gasket

(1)

 is put in place of the used 

cassettes.  Upon completion of the cleaning cycle, flush the system with WFI, or DI water 

prior to draining and discarding the cassettes.  Table 2 lists possible recommended 

cleaning solutions.  

(1)

 Spacer gaskets for flushing may be obtained by contacting 

Repligen. 

Table 2

 

Recommended Cleaning Solutions

 

Cleaning Agent

 

Cleaning Conditions 

0.5N Sodium Hydroxide

 

Contact Time = 30 – 

60 minutes

 

Temperature = 35°C (95°F) 

 

1.5% Alconox® Detergent

 

Contact Time = 30 – 

60 minute

Temperature = 40°C (104°F)

 

DISPOSAL OF USED TANGENX™ SIUS™ CASSETTES 

TangenX™ SIUS™ cassettes are removed from the holder by reversing the cassette 

installation procedure.  If the cassettes are difficult to separate from the stainless steel 

holder, a thin plastic spatula

(2)

 can be slid under the edge of the cassette and break the 

seal.  Cassettes can then be disposed of in a similar fashion to other disposable process 

equipment. 

(2)

 One cassette extractor is included with each shipment of TangenX™ 

SIUS™ cassettes. 

STORAGE OF UNUSED TANGENX™ SIUS™ CASSETTES 

Membrane cassettes must remain sealed in their original packaging prior to use to 

maintain their characteristics, integrity, and prevent microbial growth.   Below are critical 

factors to remember when storing unused TangenX™ SIUS™ cassettes: 

Recommended storage temperature: 

o

 

4°C - 25°C long term (> 7 days) 

o

 

50°C  short term (< 7 days) 

o

 

Do not freeze cassettes 

MEMBRANE OPERATING CHARACTERISTICS 

Take care to use the membrane at the lowest pressure possible while still producing 

consistent  permeate flow.  Although higher operating pressures initially improve flow 

rate, they also promote increased concentration polarization and membrane compaction, 

which ultimately limits flow.  With very low NMWL membranes, lower operating pressure 

may also reduce the retention of salts and very low molecular weight species.

 

Reviews: