IF.PUG.012     R3 

TANGENX™ PRO PD CASSETTE INSTALLATION 

1.

 

Lift the end plate off the manifold of the TangenX™ PRO PD cassette holder.

 

2.

 

Remove the protective blue sheet and rinse the EPDM gaskets with deionized 

water or WFI. Place a rinsed gasket flat against the bottom manifold; ensure 

that the holes in the gasket line up with the holes in the manifold.

 

3.

 

Using scissors carefully open the cassette bag to remove cassette.

 

WARNING: Each cassette is stored in an aqueous solution containing 

15-20% glycerin and 0.1% sodium azide, pH 7 – 10. Follow standard 

safety procedures for handling aqueous glycerin/sodium azide, 

including the use of gloves, safety goggles, and lab coat.

 

4.

 

Place the cassette into the holder flat against the gasket. Place another gasket 

on top of the cassette. Ensure that the holes in the manifold, gaskets, and 

cassette are completely aligned. If you are using multiple cassettes, continue 

the same gasket/cassette/gasket pattern, ending with a gasket between the 

last cassette and the end plate.

 

5.

 

Place the end plate on top of the last gasket of the cassette or cassette stack.

 

6.

 

Install the tie-rod spacers (if used) and washers on each bolt leaving a 

minimum of 18 mm (0.75 inch) of thread exposed on the rod.  By hand, screw 

the nut on each bolt and hand tighten evenly by alternating from one nut to 

the other.  Bolts must be further tightened to within the recommended 

torque values using a calibrated manual torque wrench. 

7.

 

Proceed to step 8 for a two bolt hardware design or skip to step 9 for a four 

bolt hardware design. 

8.

 

Using the calibrated torque wrench with a deep style socket, tighten each hex 

nut ¼ turn following the torque sequence illustrated in Figure 2 for a two bolt 

hardware.  Tighten the first nut ¼ turn, and then tighten the second nut ¼ 

turn alternating back and forth until the torque wrench "clicks".  Repeat this 

sequence until the wrench "clicks" without turning the nut.  The "click" of the 

torque wrench indicates that the nut has reached the set point torque value.  

 

9.

 

Using the calibrated torque wrench with a deep style socket, tighten each hex 

nut ¼ turn following the torque sequence illustrated in Figure 3 for a four bolt 

hardware.  In ¼ turn increments, tighten the first, second, third, and fourth 

nut alternating back and forth until the torque wrench "clicks".  Repeat this 

sequence until the wrench "clicks" without turning the nut.  The "click" of the 

torque wrench indicates that the nut has reached the set point torque value.  

 

 

CAUTION: Nuts must be tightened uniformly to avoid damaging 

the cassette.  Leakage may result from non-parallel plate 

alignment or over compression of the cassettes at one end.   

10.

 

Wait 5-10 minutes and allow the gaskets to relax before re-torqueing. 

Check each nut, per Figure 1 or 2 sequences, using the torque wrench at 

its set point torque value.    

11.

 

Re-torque as needed, to create a liquid-tight seal. 

NOTE: Torque may change during processing as the cassettes may 

compress, or as the cassettes expand or contract with 

temperature changes.  Periodically check the torque of the bolts 

and adjust torque as needed.   

 

 

FIGURE 2 

TORQUE SEQUENCE FOR TANGENX™ PRO PD CASSETTES 

(2 Bolt Design) 

 

FIGURE 3 

TORQUE SEQUENCE FOR TANGENX™ PRO PD CASSETTES 

(4 Bolt Design) 

TangenX™ PRO PD 

Cassette Holder  

COMPATIBILITY GUIDE 

REPLIGEN TANGENX™ STANDARD WARRANTY

 

Repligen Corporation

 warrants its TangenX™ products will meet their 

applicable published specifications when used in accordance with their 

applicable instructions for a period of one year from shipment of the 

products.  

REPLIGEN MAKES NO OTHER WARRANTY, EXPRESSED OR 

IMPLIED.  THERE IS NO WARRANTY OF MERCHANTABILITY OR FITNESS 

FOR A PARTICULAR PURPOSE.  

The warranty provided herein and the 

data, specifications and descriptions of Repligen TangenX™ products 

appearing in published catalogues and product literature may not be 

altered except by express written agreement signed by an officer of 

Repligen.   Representations, oral or written, which are inconsistent with 

this warranty or such publications are not authorized and if given, 

should not be relied upon. 
In the event of a breach of the foregoing warranty, Repligen’s sole 

obligation shall be to repair or replace, at its option, the applicable 

product or part thereof, provided the customer notifies Repligen 

promptly of any such breach.  If after exercising reasonable efforts, 

Repligen is unable to repair or replace the product or part, then 

Repligen shall refund to the customer all monies paid for such 

applicable product or part.  

REPLIGEN SHALL NOT BE LIABLE FOR 

CONSEQUENTIAL, INCIDENTAL, SPECIAL OR ANY OTHER DAMAGES 

RESULTING FROM ECONOMIC LOSS OR PROPERTY DAMAGE 

SUSTAINED.  

Phone 508.845.6400  |  Fax 508.845.3030 

www.repligen.com/tangenx 

[email protected] 

 

 

 

Reviews: