5.2
Value release
Enter this parameter to adjust rate of stabilization of the measurement result. Depending on the
selected option, weighing time is either shorter or longer.
1.
Enter
<P2.rEAd / 2.2.APPr>
submenu.
2.
Press
key, filter values are displayed successively one by one, where:
F_P
–
Fast and
Reliable,
PrEc
- Reliable,
FASt
–
Fast.
3.
Set respective value and press
key to confirm, next go to the home screen.
5.3
Ambient conditions
Parameter relating to ambient and environmental conditions of the workstation. Enter this
parameter and set 'nStAb' value if the ambient conditions are unfavorable (air drafts, vibrations).
4.
Enter
<P2.rEAd / 2.3.Enut>
submenu.
5.
Press
key, filter values are displayed successively one by one, where:
nStAb
–
unstable,
StAb
–
stable.
6.
Set respective value and press
key to confirm, next go to the home screen.
5.4
Autozero function
‘Autozero’ function has been designed to enable automatic control of zero indication. This
guarantees precise weighing results. There are, however, some cases where this function can be a
negative factor in the measuring process, e.g. very slow loading of the weighing pan (load adding,
e.g. pouring, filling). In such case, it is recommended to disable the function.
1.
Enter
<P2.rEAd / 2.4.Aut>
submenu.
2.
Press
key, filter values are displayed successively one by one, where:
YES
–
Function
enabled,
no
–
Function disabled.
3.
Set respective value and press
key to confirm, next go to the home screen.
5.5
Tare entering mode
The tare is entered using
and
keys combination. There are two enter modes.
1.
Enter
<P2.rEAd / 2.6.ttr>
submenu.
2.
Press
key, filter values are displayed successively one by one:
tArEH
-to enter tare
value manually,
tArnn
–
to enter tare value that is stored in scale memory.
3.
Set respective value by using
and
keys.
4.
Press
key to confirm, next go to the home screen.
5.6
Tare value memory
It is possible to store 10 tare values in scale memory.
5.6.1
Entering tare value to scale memory
1.
Enter
<P2.rEAd / 2.7.tArn>
submenu, name of tare no. 1 from tare database is displayed
(
<tArE 0>